Test of a prototype neutron spectrometer based on diamond detectors... | |
by: | |
M. Osipenko | |
F. Pompili | |
M. Ripani | |
M. Pillon | |
G. Ricco | |
B. Caiffi | |
R. Cardarelli | |
G. Verona-Rinati | |
S. Argiro | |
Thumbnail | |
Download | |
Web page | |
A prototype of neutron spectrometer based on diamond | |
detectors has been developed. This prototype consists of | |
a $^6$Li neutron converter sandwiched between two CVD | |
diamond crystals. The radiation hardness of the diamond | |
crystals makes it suitable for applications in low power | |
research reactors, while a low sensitivity to gamma rays | |
and low leakage current of the detector permit to reach | |
good energy resolution. A fast coincidence between two | |
crystals is used to reject background. The detector was | |
read out using two different electronic chains connected | |
to it by a few meters of cable. The first chain was based | |
on conventional charge-sensitive amplifiers, the other | |
used a custom fast charge amplifier developed for this | |
purpose. The prototype has been tested at various neutron | |
sources and showed its practicability. In particular, the | |
detector was calibrated in a TRIGA thermal reactor (LENA | |
laboratory, University of Pavia) with neutron fluxes of | |
$10^8$ n/cm$^2$s and at the 3 MeV D-D monochromatic | |
neutron source named FNG (ENEA, Rome) with neutron fluxes | |
of $10^6$ n/cm$^2$s. The neutron spectrum measurement was | |
performed at the TAPIRO fast research reactor (ENEA, | |
Casaccia) with fluxes of 10$^9$ n/cm$^2$s. The obtained | |
spectra were compared to Monte Carlo simulations, | |
modeling detector response with MCNP and Geant4. | |
Date Published: 2018-06-27 20:17:46 | |
Identifier: arxiv-1505.06314 | |
Item Size: 6480855 | |
Language: English | |
Media Type: texts | |
# Topics | |
Instrumentation and Detectors | |
Physics | |
Nuclear Experiment | |
# Collections | |
arxiv | |
journals | |
# Uploaded by | |
@arkiver2 | |
# Similar Items | |
View similar items | |