Equipment and procedures are described which allow the
analysis of metals in lubricating oils w/o need for
dilution or other sample pretreatment. A sample heater
has been added to a modified Babington principle
nebulizer. This nebulizer is capable of producing, from
undiluted oils, aerosols suitable for introduction into
an Inductively Coupled Plasma Atomic Emission
Spectrophotometer. Heating the samples immediately prior
to nebulization greatly increases the output of aerosol
and reduces variations in output due to differences in
sample viscosity. The type of organometallic complex used
in the preparation of standards is shown to be
unimportant if the region of the plasma which is observed
is well chosen. (Author)