Assessment of single-vessel cerebral blood velocity by phase contrast fMRI

['Xuming Chen', 'High-Field Magnetic Resonance', 'Max Planck Institute For Biological Cybernetics', 'Tübingen', 'Department Of Neurology', 'Wuhan University', 'Renmin Hospital', 'Wuhan', 'Yuanyuan Jiang', 'Athinoula A. Martinos Center For Biomedical Imaging']

Date: 2021-09

Current approaches to high-field functional MRI (fMRI) provide 2 means to map hemodynamics at the level of single vessels in the brain. One is through changes in deoxyhemoglobin in venules, i.e., blood oxygenation level–dependent (BOLD) fMRI, while the second is through changes in arteriole diameter, i.e., cerebral blood volume (CBV) fMRI. Here, we introduce cerebral blood flow–related velocity-based fMRI, denoted CBFv-fMRI, which uses high-resolution phase contrast (PC) MRI to form velocity measurements of flow. We use CBFv-fMRI in measure changes in blood velocity in single penetrating microvessels across rat parietal cortex. In contrast to the venule-dominated BOLD and arteriole-dominated CBV fMRI signals, CBFv-fMRI is comparable from both arterioles and venules. A single fMRI platform is used to map changes in blood pO 2 (BOLD), volume (CBV), and velocity (CBFv). This combined high-resolution single-vessel fMRI mapping scheme enables vessel-specific hemodynamic mapping in animal models of normal and diseased states and further has translational potential to map vascular dementia in diseased or injured human brains with ultra–high-field fMRI.

Funding: X.Y. and Y.J. were supported by NIH RF1NS113278, R01NS120594, D.K. was supported by NIH R35NS097265 and R01MH111438, X.C. was supported by the student scholarship from the Chinese Scholarship Council and Deutsche Forschungsgemeinschaft (DFG, Germany Research Foundation) grant YU 215/3-1, K.S. and R.P. were supported by DFG SCHE 658/15, SCHE 658/12. S.C. was supported by Bundesministerium fuer bildung und forschung (BMBF, Federal Ministry of Education and Research) grant 01GQ1702. The funders had no role in study design, data collection and analysis, decision to publish, or preparation of the manuscript.

Here, we report on a PC-MRI method to detect vessel-specific changes in blood velocity on the timescale of single trials. We denote this method cerebral blood flow–related velocity, abbreviated CBFv. We build on past implementations of PC-MRI [ 30 , 34 – 37 ]. Here, we utilize a small surface RF coil with high-field MRI, i.e., 14.1 T, for improved SNR. This advance allows us to map the BOLD, CBV, and CBFv-fMRI signals from individual penetrating venules and arterioles, which span 20 to 70 μm diameter, with high spatial resolution [ 22 , 38 , 39 ].

In contrast to the ASL-based orientation-specific flow measurement [ 24 – 27 ], phase contrast (PC) MRI relies on gradient-oriented dephasing of magnetized protons to map the velocity, i.e., direction and speed, of blood flow [ 28 , 29 ]. Past work with 7 T MRI showed that PC-MRI can be used to measure flow in the perforating arteries through the white matter and the lenticulostriate arteries in the basal ganglia of human brains [ 30 – 33 ]. However, the SNR was insufficient in these prior studies to map changes in flow, and thus changes in CBF, concurrent with changes in neuronal activation.

Changes in CBF may be detected by fMRI that is based on arterial spin labeling (ASL), in which water protons in a major upstream vessel are spin-polarized with an additional radio frequency (RF) B1 field [ 14 – 16 ]. The ASL-based CBF fMRI technique detects local changes in the flow of blood through brain tissue [ 17 ]. Two other fMRI-based techniques provide indirect information about changes in CBF. Blood oxygenation level–dependent (BOLD) fMRI is used to determine changes in the ratio of deoxy to oxyhemoglobin in the blood and is an indirect measure of changes in brain metabolism [ 14 , 18 , 19 ]. Cerebral blood volume (CBV) fMRI is used to measure changes in blood volume, i.e., essentially changes the diameter of arterioles, based on the use of exogenous contrast agents, e.g. MION, or the endogenous vascular-space-occupancy (VASO) mapping scheme to differentiate blood from brain tissue [ 20 – 23 ].

Cerebral blood flow (CBF) is a key hemodynamic readout coupled to neuronal dynamics and viability in normal and diseased brain states [ 1 ]. Changes in CBF may be monitored directly within individual blood vessels through the use of optical-based particle tracking techniques [ 2 ]. A variety of imaging methods have been developed to measure CBF across multiple spatial scales, from capillary beds up through brain-wide vascular networks in animal brains. These include in vivo multiphoton microscopy [ 3 ], optical coherence tomography [ 4 ], optoacoustic imaging [ 5 ], and laser doppler and speckle imaging [ 6 , 7 ]. In particular, doppler based functional ultrasound imaging provides a unique advantage to detect the CBF in the brain with a high spatiotemporal resolution, which can be readily applied for awake animal imaging [ 8 – 10 ]. However, the spectrum-specific signal transmission in ultrasound methods cannot effectively pass the skull of animals without significant degradation of the signal-to-noise ratio (SNR). Typically, a craniotomy or a procedure to thin the skull is needed to detect the hemodynamic signal [ 2 ]. While current techniques support transcranial imaging into the superficial layers of the cortex [ 11 – 13 ], only functional MRI (fMRI) provides a noninvasive approach for measuring hemodynamic signals throughout the brain.

The stimulus-evoked responses of all CBV, BOLD, and CBFv-fMRI signals were studied with an on/off block design ( Fig 3B–3E ). Group analysis shows that the positive BOLD signal from venule voxels is significantly higher than the arteriole-specific BOLD signal ( Fig 3C ). In contrast, the arteriole dilation leads to an earlier CBV-weighted negative fMRI signal, which is much stronger and faster than the signal from passive venule dilation ( Fig 3C ), as expected [ 45 – 47 ]. Group analysis shows the similar temporal dynamics of CBFv changes, but with different signs in arterioles and venules ( Fig 3D ). In contrast to the venule-dominated BOLD and arteriole-dominated CBV responses ( Fig 3C and 3F ), the percentage changes in CBFv-fMRI is statistically indistinguishable between arterioles (3.12 ± 0.87%) and venules (3.60 ± 0.80%) ( Fig 3E and 3F ). The voxel-wise hemodynamic changes of BOLD, CBV, and CBFv are illustrated in S3 Fig and S1 Movie .

(A) Different MRI measurement strategies on the same 2D slice. From left to right: (i) the A–V map defines arterioles as bright dots and venules as dark dots; (ii) the evoked bSSFP-based BOLD fMRI map, within a green subregion, on top of the A–V map; (iii) the PC-MRI map of baseline CBFv; (iv) the CBFv-fMRI map on top of the A–V map with an increased flow velocity corresponding to brighter voxels for venules and darker voxels for arterioles; and (v) evoked bSSFP-based CBV-fMRI map on top of the A–V map. (B) The time courses of evoked bSSFP-BOLD and CBV-fMRI with the block design paradigm from venules and arterioles shown in panel A. Forepaw stimulation pulse of 330 μs in width and 1 mA in amplitude delivered at 3 Hz for 10 seconds. (C) Averaged time courses of the fractional change for evoked BOLD and CBV signals from venule and arteriole ROIs of different rats (mean ± SEM, the green bar shows stimulation duration). (D) The time courses of the evoked CBFv changes from the arteriole and venule ROIs show increased velocity from both arterioles and venules with the block design, 10-second duration stimulation paradigm. (E) The averaged time courses of the evoked CBFv changes show the velocity increase from both arteriole and venule ROIs with the block design stimulation paradigm from 5 rats (mean ± SEM, the green bar shows stimulation duration). (F) Bar graph shows that peak BOLD changes of venule are significantly higher than those of arteriole (5 rats, p = 0.009), while the peak CBV changes of arteriole are significantly higher than those of venule (3 rats, p = 0.028). In contrast, the peak CBFv changes are only slightly, but not significantly, higher in venules than arteriole (5 rats, p = 0.063). The data underlying this figure can be found in S3 Data . A–V, arteriole–venule; BOLD, blood oxygenation level–dependent; bSSFP, balanced steady-state free precession; CBFv, cerebral blood flow–related velocity; CBV, cerebral blood volume; ROI, region of interest.

We contrasted the complementary capabilities of PC-based CBF-fMRI against the signals observed with the balanced steady-state free precession (bSSFP)-based single-vessel BOLD- and CBV-fMRI mapping method [ 38 ] ( Fig 3 ). We first created an A–V map through the infragranular layers of the forepaw region of the primary somatosensory cortex ( Fig 3Ai ), followed by 2D-bSSFP to detect stimulus-induced changes in the single-vessel BOLD fMRI signal ( Fig 3Aii ). We next performed single-vessel PC-MRI flow velocity measurements to measure baseline flow in penetrating arterioles and venules, using 100 × 100 μm 2 in-plane resolution, a sampling rate of 4-second repetition time (TR) per image and the same geometry as the 2D-bSSFP method ( Fig 3Aiii ). Changes in CBFv upon stimulation overlapped with individual penetrating vessels in the A–V map ( Fig 3Aiv ). Lastly, we performed 2D-bSSFP for single-vessel CBV-fMRI mapping by intravenous injection of iron particles into the blood in the same animals ( Fig 3Av ). The BOLD fMRI signal is primarily detected from individual penetrating venules, while the CBV-weighted signal is mainly located at the individual penetrating arterioles (dark dots in Fig 3Aii with bright dots in Fig 3Av ). In contrast, the CBFv-fMRI signal is observed in both the penetrating arterioles and venules ( Fig 3Aiv ).

We first acquired the single-vessel A–V map by aligning a 500-μm thick 2D MRI slice perpendicular to penetrating vessels through layer V of one hemisphere ( Fig 2A and 2B ). We designed the pulse sequence for PC-MRI to achieve the same slice geometry of the A–V map so that the CBF deduced from PC-MRI signals could be overlaid with individual penetrating arterioles and venules in the single-vessel flow velocity map ( Fig 2A and 2C ). The arteriole blood flows into the cortex, while the venule blood flows outward, which determines the sign of the flow velocity. Vessel-specific velocities were plotted as a function of the normalized signal intensity in the A–V map and corroborated our ability to determine flow velocity specific to arterioles and venules ( Fig 2C and 2D ). The measured flow velocities range from 1 to 10 mm/s, as previously measured with optical methods [ 43 , 44 ]. To probe the reliability of the single-vessel MR-based flow velocity method, we compared the velocities detected by PC-MRI methods with different TEs and flip angles and observed comparable results across a range of parameters ( S2 Fig ). It should be noted that altered vessel velocities detected from the same animal depend on the vessel sizes and orientation angles relative to the 2D slice. Lastly, the large variability of blood velocity across different animals could be caused by the varied physiological states of animals under anesthesia, as well as by degraded gradient performance during high duty cycle PC-MRI; see optimization step in Methods. All told, these data demonstrate the feasibility of in vivo PC-based blood velocity mapping from individual penetrating arteriole and venules.

(A) Representative 2D MGE slices (yellow boxes) from a deep layer of the primary forepaw somatosensory cortex (first frame) at different TEs, as indicated. (B) The 2D A–V map (yellow box) derived from the images with different TEs in panel A, arterioles, and venules appear as bright and dark voxels, respectively. The expanded views (red boxes) show individual venules, i.e., black voxels marked in blue, and arterioles, i.e., white voxels marked in red. (C) The vectorized flow velocity map (blue box) from the same 2D MGE slice in panel B. The expanded views (green boxes) show the individual venules, i.e., white dots with positive velocity, and arterioles, i.e., black dots with negative velocity. Note that 2 bright dots are caused by the “overflowed” velocity beyond the maximal velocity, i.e., the Venc parameter, defined in the PC-MRI sequence, which could be not correctly estimated. (D) Scatter plot of the flow velocities from individual arterioles and venules as the function of the normalized signal intensities of each vessel in the A–V map of panel B, data from 11 rats as indicated. Insert shows the blood flow direction of arterioles and venules in the forepaw somatosensory cortical region. The lower panel shows the histogram of the blood velocity distribution across arterioles and venules, as well as the bar graph to show the mean velocity. The data underlying this figure can be found in S2 Data . A–V, arteriole–venule; CBFv, cerebral blood flow–related velocity; MGE, multi-gradient echo; PC, phase contrast; ROI, region of interest; Venc, velocity encoding; TE, echo time.

We implemented the high-resolution PC-MRI with 14.1 T MR scanning for in vivo measurement of blood flow from individual penetrating arterioles and venules. We collected data throughout the infragranular cortex, i.e., layer V, of anesthetized rats. A surface RF transceiver coil with 6-mm diameter was constructed and attached to the rat skull ( S1 Fig ) to improve the SNR of PC-MRI images as well as multi-gradient echo (MGE) images used for arteriole–venule (A–V) mapping [ 22 , 38 ]. This coil was essential for the high-resolution mapping with a fast sampling rate of the single-vessel flow velocity over a complete hemisphere of the rat brain ( Figs 2A and S1E ).

(A) Schematic drawing of the phantom experimental flow chamber in the 14.1 T scanner. An expanded image (red box) shows the circulatory system constructed of capillary tubes. A representative FLASH MRI image (blue box), 500 μm in thickness, shows the capillary positions. ROIs T1 and T2, contoured in purple, indicate the upward flow. ROIs T3 and T4, in orange contour, indicate the downward flow. The green contour indicates the stagnant fluid. (B) Representative images with different flow velocity in the capillaries T1 to T4 in panel A. TE = 5.0 ms for all panels. (C) The plot of flow velocity estimates from the 5 ROIs with different TEs, as marked, and different pump rates, as indicated and marked in panel B. The dotted lines correspond to a linear fitting for velocity measurements of different ROIs. The data underlying this figure can be found in S1 Data . FLASH, fast low angle shot; PC, phase contrast; ROI, region of interest; TE, echo time.

For calibration, we constructed an in vitro capillary tubing circulatory system to mimic penetrating vessels, with flow rates from 1 to 10 mm/s ( Fig 1A ). A 2D PC-MRI slice is aligned perpendicular to the capillary tubing ( Fig 1A and 1B ) and provides a voxel-specific measurement of the flow velocity through 2 tubes with the upward flow (positive sign, bright dots in Fig 1B ) and 2 tubes with the downward flow (negative sign, dark dots in Fig 1B ), as well as a control tube. We observe a monotonic and near linear relation between the velocity measured by PC-MRI and the true velocity: V meas = (0.67 ± 0.01) Vpump + (0.02 ± 0.11) mm/s at echo time (TE) = 5.0 ms ( Fig 1C ). The small offset could be caused by eddy current effects and other gradient-related scaling errors of the PC-MRI sequence [ 40 – 42 ]. We further observe that the measured velocities are relatively insensitive to the value of TE ( Fig 1C ).

Discussion

Despite the plethera of existing tools developed for CBF measurements in both animal and human brains, it remains challenging to noninvasively detect the flow dynamics of intracortical microvessels. Here, we optimized PC-MRI to map the vectorized single-vessel flow velocity of penetrating arterioles and venules and further developed the single-vessel CBFv-fMRI technique to directly measure flow velocity in rat brains. In combination with previously established single-vessel BOLD- and CBV-fMRI methods, the PC-based single-vessel CBFv-fMRI method provides complementary information to map vessel-specific hemodynamic responses with high-resolution fMRI.

In contrast to the conventional ASL methods, PC-based MRI mapping allows arterioles and venules to be distinguished for simultaneous velocity measurements through a 2D plane. Also, ASL has less vascular specificity because water exchange through the blood–brain barrier of capillary beds increases the weighting of the ASL-based flow signal for parenchyma voxels [48,49]. Furthermore, there is significant variability in the transit time to flow from arterioles to venules through the capillary bed [50], which complicates the distinction of arterioles and venules by simple ASL-based CBF mapping. We detected the velocity from penetrating microvessels in the deep cortical layers with PC-MRI, with velocity values from 1 to 10 mm/s (Fig 2). It is noteworthy that the PC-based vessel velocity measurement is based on measuring water protons in blood but not limited to the flow of red blood cells. Still, the PC-based velocity from microvessels matches well with the previous optical measurement [2,43,44]. We conclude that high-resolution PC-MRI is ideal for noninvasive single-vessel CBFv-fMRI mapping.

The noninvasive measurement of microvessel blood velocity changes coupled to neuronal activation is a critical step to elucidate the neurovascular coupling mechanism underlying a variety of brain disorders in animal models. Existing methods, e.g., multiphoton microscopy [3,51,52] or doppler based ultrasound imaging [8–10], enable the detection of red blood cell velocity or CBF/CBV from microvessels with ultra-high resolution. Yet ultrasound requires a craniotomy or thinned skull [53] for effective spectrum-specific signal transmission through deep cortical layers or subcortical regions in rats and other animals. Such a surgical procedure has been well documented to induce perturbation of the physiological microenvironment in the brain [54–56], which potentially confounds the disease-related functional CBF measurements in animal brains. In contrast, conventional MRI methods, including PC-MRI, mainly detect blood flow or velocity in large vessels and seldom identify the functional CBFv of penetrating microvessels in the brain. By implementing a local RF coil with the high-field MR scanner, our work showed the feasibility to map the velocity changes from individual penetrating arterioles and venules with diameter 20 to 70 μm [22,38].

The novel applications of PC-based CBFv-fMRI can be divided into 2 different directions. First, the PC-based CBFv-fMRI provides a possible scheme to be combined with single-vessel BOLD and CBV fMRI signals for quantitative assessment of the cerebral metabolic rate of oxygen utilization in relation to flow [57]. These vessel-specific hemodynamic measurements can then be used to potentially validate the calibrated BOLD model [58] in future studies. Secondly, we now have a means to test the resting-state single-vessel correlation patterns of CBFv signals from penetrating arterioles and venules (S4 Fig). In contrast to the T2*-weighted resting-state readout of BOLD or CBV signal fluctuation [38], which amplifies the contrast-to-noise ratio (CNR) from the vessel-specific oxygenation or vessel diameter changes, CBFv signal fluctuation from penetrating microvessels provides a direct measurement of blood velocity changes across a large field of view. The measurement of PC-based CBFv signal fluctuation from microvessels can be further combined with single-vessel CBV-fMRI to study the vascular dynamic mechanism that underlies vasomotion-mediated perivascular clearance in animal models with Alzheimer disease (AD) [59,60]. In particular, both vasomotion-based ateriole diameter changes, i.e., the CBV signal fluctuation, and corresponding blood velocity changes, i.e., CBFv signal fluctuation, can be mapped across hippocampal penetrating vessels. This may be further combined with simultaneous fiber photometry base fluorescent recordings from genetically encoded biosensors [39] to identify the pathological vasodynamics in transgenic AD animals.

The accurate measurement of PC-based CBFv from microvessels relies on multiple factors. First, a remaining complication with PC-MRI mapping is the presence of small offsets in velocity as shown in our phantom capillary tubing studies with circulating flow under different conditions. The phase-dependent velocity encoding depends on the quality of the magnetic field gradients, and mismatched eddy currents of multiple gradients with opposite polarities, as well as the nonlinear and distorted gradient fields, could contribute to distortions in gradients [40–42]. We have described a list of parameters that should be optimized when implementing high-resolution PC-based CBFv measurements (Methods). Also, the high-resolution PC-MRI method is a high duty cycle sequence and slight heating of the gradient coil during scanning may alter the gradient performance, consistent with the baseline drift of the CBFv-fMRI signal in the first 5 minutes of scanning (Fig 3D). Nevertheless, it should be noted that the percentage velocity changes from individual arterioles and venules can be readily detected with the PC-based CBFv-fMRI measurement regardless of the gradient heating–related baseline drift. Another factor that contributes to the phase-dependent velocity error originates from the limited spatial resolution of PC-MRI images in comparison to the diameters of small vessels, i.e., the partial volume effect [36], although corrections are possible [30,37].

The present work shows the feasibility of PC-based CBFv-fMRI in rat brains using a ultra–high-field strength MR scanner (14 T). It should be noted that the high-resolution PC-based CBFv measurement can also be applied with a 9.4 T scanner to detect the microvessel blood velocity at a similar scale to 14 T measurement (S5 Fig). This supports a broader usage of noninvasive animal CBFv mapping. The translational potential of the PC-based CBFv-fMRI remains to be investigated. Previously, we detected low-frequency fluctuation of single-vessel resting state BOLD fMRI signals (TR = 1 second) from individual sulcus veins and arteries in the occipital lobe of the human brain [38]. The T2*-weighted BOLD resting state fMRI signal fluctuation from individual vessels can be separated from noise artifacts due to pulsation or other motion effects. In contrast, pulsation could be a significant confounding issue for PC-based blood velocity mapping from individual vessels [61,62]. To better differentiate the pulsational contribution to CBFv signal fluctuation given different frequency ranges, we will need to increase the sampling rate by implementing phased-array surface coils for focal field of view measurement and an accelerated PC-MRI sequence.

[END]

[1] Url: https://journals.plos.org/plosbiology/article?id=10.1371/journal.pbio.3000923

(C) Plos One. "Accelerating the publication of peer-reviewed science."
Licensed under Creative Commons Attribution (CC BY 4.0)
URL: https://creativecommons.org/licenses/by/4.0/


via Magical.Fish Gopher News Feeds:
gopher://magical.fish/1/feeds/news/plosone/