TITLE 32:  ENERGY
                  CHAPTER II:  DEPARTMENT OF NUCLEAR SAFETY
                     SUBCHAPTER b:  RADIATION PROTECTION

                                  PART 360
        USE OF X-RAYS IN THE HEALING ARTS INCLUDING MEDICAL, DENTAL,
                      PODIATRY, AND VETERINARY MEDICINE


Section
360.10      Scope
360.20      Definitions
360.30      General Requirements and Administrative Controls
360.40      General Equipment and Operation Requirements for Diagnostic X-Ray
           Systems
360.50      Fluoroscopic Systems
360.60      Stationary Radiographic Systems Other Than Fluoroscopic, Dental
           Intraoral, Veterinary, or Systems Used Solely for Mammography
360.70      Mobile/Portable Radiographic Systems Other Than Systems Used
           Solely for Mammography
360.71      Requirements for Facilities Performing Mammography
360.80      Photofluorographic Systems
360.90      Intraoral Dental Radiographic Systems
360.100     Veterinary Radiographic Systems
360.110     Therapeutic X-Ray Installations
360.120     Special Requirements for X-Ray Therapy Equipment Operated at
           Potential of Fifty (50) kVp and Below
360.APPENDIX A    Medical Radiographic Exposure Limits
360.APPENDIX B    Mammography Dose Limit
360.ILLUSTRATION A  Thimble and Pancake Chamber - Radiation Measuring Devices
360.ILLUSTRATION B  Mammography Dose Evaluation Graph
360.TABLE A  Filtration Required as a Function of Operating kVp (Repealed)
360.TABLE B  Half-Value Layer as a Function of Tube Potential
360.TABLE C  Entrance Exposure Limits Per Intraoral Bitewing Film

AUTHORITY:  Implementing and authorized by the Radiation Protection Act of
1990 (P.A. 86-1341, effective September 7, 1990)

SOURCE:  Filed April 20, 1974 by the Department of Public Health; old rules
repealed, new rules adopted at 4 Ill. Reg. 25, p. 157, effective July 1, 1980;
transferred to the Department of Nuclear Safety by P.A. 81-1516, effective
December 3, 1980; codified at 7 Ill. Reg. 16406; amended at 10 Ill. Reg.
13271, effective July 28, 1986; amended at 13 Ill. Reg. 803, effective
April 1, 1989; amended at 15 Ill. Reg. 6180, effective April 16, 1991.

Section 360.10  Scope

     a)    This Part establishes requirements for use of x-ray producing
           devices in the healing arts by a practitioner licensed to practice
           a treatment of human ailments by virtue of the Medical Practice
           Act of 1987 (Ill. Rev. Stat. 1987, ch. 111, pars. 4400-1 et seq.),
           the Illinois Dental Practice Act (Ill. Rev. Stat. 1987, ch. 111,
           pars. 2301 et seq.), or the Podiatric Medical Practice Act of 1987
           (Ill. Rev. Stat. 1987, ch. 111, pars. 4801 et seq.), or by a
           medical radiographer or radiation therapy technologist accredited
           in accordance with the provision of 32 Ill. Adm. Code 401.100 or
           an individual exempt from the provisions of 32 Ill. Adm. Code 401,
           by Section 401.30 of that Part, acting under the supervision,
           prescription or direction of such licensed person or the non-human
           use of x-ray by veterinarians by virtue of the Veterinary Medicine
           and Surgery Practice Act of 1983 (Ill. Rev. Stat. 1987, ch. 111,
           pars. 7001 et seq.).  The provisions of this Part are in addition
           to, and not in substitution for, other applicable provisions of 32
           Ill. Adm. Code 310, 320, 340, 400, and 410.

     b)    It is recognized that some installations and equipment designed
           before the adoption of this Part, coupled with conditions of use,
           may be adequate to achieve minimum exposures.  Request for
           exemption from some provisions of this Part will be considered in
           accordance with 32 Ill. Adm. Code 310.30(a).

(Source:  Amended at 13 Ill. Reg. 803, effective April 1, 1989)

Section 360.20  Definitions

As used in this Part, the following definitions apply:

           "Accelerator" means any therapeutic machine capable of producing a
           useful beam of x-rays or charged particles with energies greater
           than 500 keV.

           "Added filtration" means the effect of the material (filter) added
           to the inherent filtration.

           "Aluminum equivalent" means the thickness of type 1100 aluminum
           alloy affording the same attenuation, under specified conditions,
           as the material in question.  The nominal chemical composition of
           type 1100 aluminum alloy is 99.00 percent minimum aluminum, 0.12
           percent copper.

           "Automatic exposure control" means a device which automatically
           controls 1 or more technique factors in order to obtain at a
           preselected location(s) a required quantity of radiation (see
           "Phototimer").

           "Barrier" (see "Protective barrier").

           "Beam axis" means a line from the source through the center of the
           x-ray field.

           "Beam-limiting device" means a device which provides a means to
           restrict the dimensions of the x-ray field (see "Collimator",
           "Diaphragm", and "Shutter").

           "Certified system" means an x-ray system which is subject to
           regulations promulgated under Public Law 90-602, the Radiation
           Control for Health and Safety Act of 1968 (42 U.S.C.A. 263(b) et
           seq.), 21 CFR 1000.3 et seq., in effect as of April 1, 1988,
           exclusive of subsequent amendments or editions.  A copy of this
           document is available for public inspection at the Illinois
           Department of Nuclear Safety (Department), 1035 Outer Park Drive,
           Springfield, Illinois.

           "Collimator" means a device or mechanism by which the x-ray beam
           is restricted in size (see "Beam-limiting device").

           "Contact therapy system" means an x-ray system used for therapy
           with the x-ray tube port placed in contact with or within 5
           centimeters of the surface being treated.

           "Control panel" means that part or parts of the x-ray system upon
           which are mounted the switches, knobs, pushbuttons, and other
           hardware necessary for setting the technique factors prior to
           initiating an x-ray exposure.

           "Dead-man switch" means a switch so constructed that a circuit-
           closing contact can be maintained only by continuous pressure on
           the switch by the operator.

           "Diagnostic source assembly" means an x-ray tube housing assembly,
           designed for use in diagnostic x-ray applications, with a beam
           limiting device attached.

           "Diagnostic-type protective tube housing" means an x-ray tube
           housing constructed so that when a beam limiting device is
           attached, the leakage radiation measured at a distance of 1 meter
           from the source cannot exceed 100 mR in 1 hour when the tube is
           operated at its maximum continuous rated current for the maximum
           rated tube potential.

           "Diaphragm" means a device or mechanism by which the x-ray beam is
           restricted in size (see "Beam-limiting device").

           "Filter" means material placed in the useful beam to absorb,
           preferentially, radiations based on energy level (see
           "Filtration" and "Inherent filtration").

           "Filtration" means the act of preferentially absorbing radiation
           with filters or inherent filtration (see "Filter" and "Inherent
           filtration").

           "General purpose radiographic x-ray system" means any radiographic
           x-ray system which, by design, is not limited to radiographic
           examination of specific anatomical regions.

           "Gonad shield" means a protective device for the testes or ovaries
           which provides a minimum of 0.50 mm lead equivalent protection.

           "Half-value layer (HVL)" means the thickness of a specified
           material that attenuates the beam of radiation to an extent such
           that the exposure rate is reduced to one-half of its original
           value.

           AGENCY NOTE:  The contribution of all scattered radiation, other
           than any that might be present initially in the beam concerned, is
           minimized.

           "Healing arts screening" means the examination of human beings
           using x-ray machines for the detection or evaluation of potential
           diseases when such examinations are not specifically ordered by a
           licensed practitioner of the healing arts legally authorized to
           prescribe such x-ray examinations for the purpose of diagnosis or
           treatment.  However, healing arts screening does not include
           mammography on self-referred patients.

           "Image intensifier" means a device, installed in a housing, which
           converts an x-ray pattern into a corresponding light image.

           "Image receptor" means any device, such as a fluorescent screen or
           radiographic film, which transforms incident x-ray photons either
           into a visible image or into another form which can be made into a
           visible image by further transformations.

           "Inherent filtration" means the filtration permanently in the
           useful beam; it includes the effect of the x-ray tube window and
           any permanent tube or source enclosure (see "Filter" and
           Filtration").

           "Interlock" means a device arranged or connected such that the
           occurrence of an event or condition is required before a second
           event or condition can occur or continue to occur.

           "Kilovolts peak (kVp)" means the crest value, in kilovolts, of the
           electric potential applied to the x-ray tube between the cathode
           and anode of a pulsating electric potential generator.

           "Lead equivalent" means the thickness of lead affording the same
           attenuation, under specified conditions, as the material in
           question.

           "Leakage radiation" means all radiation emanating from the
           diagnostic source assembly except for:

                 The useful beam; and

                 The radiation produced when the exposure switch or timer is
                 not activated.

           "Leakage technique factors" means the technique factors used to
           measure leakage radiation from the diagnostic source assembly.
           They are defined as follows:

                 For capacitor energy storage equipment, the maximum-rated
                 peak tube potential and the maximum-rated number of
                 exposures in an hour for operation at the maximum-rated peak
                 tube potential with the quantity of charge per exposure
                 being 10 millicoulombs, i.e. 10 milliampere seconds, or the
                 minimum obtainable from the unit, whichever is larger.

                 For field emission equipment rated for pulsed operation, the
                 maximum-rated peak tube potential and the maximum-rated
                 number of x-ray pulses in an hour for operation at the
                 maximum-rated peak tube potential.

                 For all other equipment, the maximum-rated peak tube
                 potential and the maximum-rated continuous tube current for
                 the maximum-rated peak tube potential.

           "Light field" means that area of the intersection of the light
           beam from the beam-limiting device and any one of the sets of
           planes parallel to and including the plane of the image receptor.
           The edge of the light field is defined as the locus of points at
           which the illumination is 25 percent of that at the center of the
           light field.

           "Mammography" means radiography of the breast for the purpose of
           enabling a physician to determine the presence, size, location and
           extent of cancerous or potentially cancerous tissue in the breast.

           "Mammography dosimetry test phantom" means a phantom for
           determining the mean glandular breast dose.  For automatic
           exposure control mammography systems it shall be any phantom
           material that is equivalent to a nominal 4.5 centimeter compressed
           breast of average density (i.e., 50 percent adipose and 50 percent
           glandular tissue).

           "Mammography System" means an x-ray system that is used to perform
           mammography.

           "Medical radiographer" means a person other than a licensed
           practitioner, accredited in accordance with the provisions of 32
           Ill. Adm. Code 401, or an individual exempt from the provisions of
           32 Ill. Adm. Code 401, who performs medical radiation procedures
           and applies x-radiation, to any part of the human body, for
           diagnostic purposes while under the supervision of a licensed
           practitioner.

           "Mobile equipment" (see "X-ray equipment").

           "Non-certified system" means an x-ray system which is not subject
           to regulations promulgated under Public Law 90-602, the Radiation
           Control for Health and Safety Act of 1968 (see "Certified
           system").

           "Personnel monitoring" means the determination of radiation
           exposure to a person.  Devices used for this purpose may include,
           but are not limited to, film badges, pocket dosimeters, and
           thermoluminescent dosimeters worn by the individual.

           "Phototimer" means a method for controlling radiation exposures to
           image receptors by the amount of radiation which reaches a radia-
           ion monitoring device(s).  The radiation monitoring device(s) is
           part of an electronic circuit which controls the duration of time
           the tube is activated (see "Automatic exposure control").

           Portable equipment" (see "X-ray equipment").

           "Position indicating device" means a device on intraoral dental
           x-ray equipment used to indicate the beam position and to
           establish a definite source-skin distance.

           "Positive beam limitation" means a beam-limiting device which
           will, at the source-image receptor distance for which the device
           is designed, either cause automatic adjustment of the x-ray field
           in the plane of the image receptor to the image receptor size
           within 5 seconds after insertion of the image receptor or, if
           adjustment is accomplished automatically in a time interval
           greater than 5 seconds or is manual, prevent production of x-rays
           until such adjustment is completed.  For SIDs at which the device
           is not intended to operate, the device prevents the production of
           x-rays.

           "Primary protective barrier" (see "Protective barrier").

           "Protective apron" means an apron of radiation absorbing
           materials, at least 0.25 mm lead equivalent, used to reduce
           exposure from stray radiation (see "Stray radiation").

           "Protective barrier" means a barrier of radiation absorbing
           material(s) used to reduce radiation exposure.  The types of
           protective barriers are as follows:

                 "Primary protective barrier" means the material, excluding
                 filters, placed in the useful beam to reduce the radiation
                 exposure.

                 "Secondary protective barrier" means a barrier sufficient to
                 attenuate the stray radiation to the required degree (see
                 "Stray radiation").

           "Protective glove" means a glove made of radiation absorbing
           materials, at least 0.25 mm lead equivalent, used to reduce
           exposure from stray radiation (see "Stray radiation").

           "Qualified expert" means an individual who has demonstrated to the
           satisfaction of the Department that he or she possesses the
           knowledge and training to measure ionizing radiation, to evaluate
           safety techniques, and to advise regarding radiation protection
           needs.  Satisfactory demonstration of such knowledge and training
           should include certification by a nationally recognized
           credentialing entity in the field of radiation protection.
           "Radiation therapy simulation system" means a radiographic/
           fluoroscopic x-ray system used exclusively for localizing the
           volume to be exposed during radiation therapy and confirming the
           position and size of the therapeutic irradiation field.

           "Radiologist" means a physician or veterinarian who is either:

                 Certified by the American Board of Radiology in diagnostic
                 radiology or general radiology;

                 Certified by the American Osteopathic Board of Radiology;

                 Certified by the American Chiropractic Board of Radiology;
                 or

                 Certified by the American College of Veterinary Radiology;
                 or

                 Eligible for certification by any College or Board
                 identified above.

           "Scatter radiation" means radiation that, during passage through
           matter, has been deviated in direction.

           "Secondary protective barrier" (see "Protective barrier").

           "Shutter" means an adjustable beam-limiting or attenuating device,
           usually made of lead, fixed to an x-ray tube housing to intercept
           or collimate the useful beam (see "Beam-limiting device").

           "SID" means source-image receptor distance (see "Source-image
           receptor distance").

           "Source" means the focal spot of the x-ray tube.

           "Source-image receptor distance" means the distance from the
           source to the center of the input surface of the image receptor.

           "Source-skin distance (SSD)" means the distance measured along the
           central ray from the center of the front surface of the x-ray
           focal spot to the surface of the irradiated object.

           "Special purpose x-ray system" means any radiographic x-ray system
           which, by design, is limited to radiographic examination of a
           specific anatomical region.

           "Spot film" means a radiograph which is made during a fluoroscopic
           examination to permanently record conditions which exist during
           that fluoroscopic procedure.

           "Stationary equipment" (see "X-ray equipment").

           "Stray radiation" means the sum of leakage and scatter radiation.

           "Technique factors" means the electrical potential (kilovolts),
           current (milliamperes), exposure time parameters (seconds or
           pulses) or a combination thereof, selectable at the control panel
           of an x-ray system (see "Control panel").

           "Useful beam" means the radiation which passes through the tube
           housing port and the aperture of the beam-limiting device when the
           exposure switch or timer is activated.

           "X-ray equipment" means an x-ray system, sub-system, or component
           thereof.  Types of x-ray equipment are as follows:

                 "Mobile x-ray equipment" means x-ray equipment mounted on a
                 permanent base with wheels and/or casters for moving while
                 completely assembled.

                 "Portable x-ray equipment" means x-ray equipment designed to
                 be hand-carried.

                 "Stationary x-ray equipment" means x-ray equipment which is
                 installed in a fixed location.

           "X-ray field" means, for diagnostic purposes, that area of the
           intersection of the useful beam and any 1 of the set of planes
           parallel to and including the plane of the image receptor.  The
           edge of the x-ray field is defined as the locus of points at which
           the exposure is 25 percent of that at the center of the x-ray
           field.

           "X-ray system" means an assemblage of components for the
           controlled production of x-rays.  It includes minimally an x-ray
           high-voltage generator, an x-ray control panel, an x-ray tube
           housing assembly, a beam-limiting device, and the necessary
           supporting structures.  Additional components which function with
           the system are considered integral parts of the system.

(Source:  Amended at 15 Ill. Reg. 6180, effective April 16, 1991)

Section 360.30  General Requirements and Administrative Controls

The requirements in this Section apply to all uses of x-rays in veterinary
medicine and to all uses of x-rays in the Healing Arts including the use of
x-rays for both diagnostic and therapeutic purposes.  Additional requirements
for all diagnostic x-ray systems are in Section 360.40 and specific equipment
application classes are contained in Sections 360.50 through 360.100.  For
therapeutic x-ray systems also see Sections 360.110 and 360.120.

     a)    Registrant - The registrant shall:

           1)    Direct the operation of the x-ray system(s);

           2)    Register with the Department, in accordance with the
                 provisions of 32 Ill. Adm. Code 320, all x-ray equipment
                 which is used at the facility;

           3)    Register with the Department, in accordance with the
                 provisions of 32 Ill. Adm. Code 320, all portable or mobile
                 x-ray equipment used by the Registrant;

           4)    Submit an application for inspection of radiation machines
                 to the Department in accordance with 32 Ill. Adm. Code 410
                 and, if the inspection is performed by a qualified
                 nondepartment inspector, submit a copy of the radiation
                 inspection report to the Department;

           5)    Permit operation of the x-ray system(s) only by individuals
                 who are licensed in accordance with State law (See Section
                 360.10(a)), or who are accredited by the Department or are
                 exempt from such requirements in accordance with the
                 provisions of 32 Ill. Adm. Code 401;

           6)    Inform all individuals who work in activities pursuant to
                 the operator's registration of their rights in accordance
                 with the provisions of 32 Ill. Adm. Code 400; and

           7)    Maintain records showing the receipt, transfer, use,
                 storage, and disposal of all sources of radiation in
                 accordance with the provisions of 32 Ill. Adm. Code 310 and
                 320.

     b)    Shielding - Each installation shall be provided with such primary
           barriers and/or secondary barriers as are necessary to assure
           compliance with the provisions of 32 Ill. Adm. Code 340.1010,
           340.1040 and 340.1050.

     c)    An x-ray system which does not meet the provisions of this Part
           shall not be operated for diagnostic or therapeutic purposes if so
           ordered by the Director.

     d)    If an x-ray system is identified as not being in compliance with
           the provisions of this Part and if that system is accessible for
           use, it shall be rendered inoperable (i.e. dismantle the x-ray
           source from the source support assembly) if so ordered by the
           Director.

     e)    Unauthorized Exposure - Individuals shall not be exposed to the
           useful beam except for healing arts purposes and only when such
           exposure has been authorized by a licensed practitioner of the
           healing arts.  This provision specifically prohibits deliberate
           exposure for the following purposes:

           1)    Exposure of individuals for training, demonstration, or
                 other non-healing arts purposes.

           2)    Exposure of individuals for the purpose of "healing arts
                 screening" (see Section 360.20).

     f)    Personnel Monitoring and Reporting Requirements -  All persons who
           are associated with the operation of an x-ray system are subject
           to the occupational exposure limits and the requirements for the
           determination of the doses which are contained in 32 Ill. Adm.
           Code 340.1010, 340.1020, 340.1040, 340.2010, and 340.2020 and the
           reporting requirements as stated in 32 Ill. Adm. Code 340.4010
           through 340.4080.

     g)    The registrant shall comply with the requirements of the
           Department's rules entitled, Notices, Instructions and Reports to
           Workers; Inspections, 32 Ill. Adm. Code 400.

     h)    Maintenance Records and Associated Information - The registrant
           shall maintain, for a period of at least 3 inspection cycles (see
           32 Ill. Adm. Code 410.60(d)), the following information for each
           x-ray system for inspection by the Department:

           1)    Separate records of maintenance and modifications performed
                 on each x-ray system with the name(s) of the individual(s)
                 who performed such services and the date(s) performed.

           2)    A copy of all correspondence with the Department regarding
                 the registrant's x-ray program.

     i)    Staff Qualifications - The registrant shall maintain for review by
           the Department:

           1)    A current staffing plan indicating the names of all
                 individuals responsible for operating x-ray equipment and
                 the scope of their duties at the facility.

           2)    Current certificates of accreditation (clear, legible copies
                 are acceptable), issued by the Department in accordance with
                 the provisions of 32 Ill. Adm. Code 401, for all individuals
                 who are required to be so accredited.

     j)    Radiation Safety Program - The registrant shall provide for annual
           in-service training in radiation safety for individuals (excluding
           licensed practitioners) that apply ionizing radiation at the
           facility, to ensure their awareness of the registrant's radiation
           safety practices and policies.

           1)    The in-service training must include the following topics:

                 A)    Operating and emergency procedures for the radiation
                       machine(s);

                 B)    Use of personnel and patient protective devices;

                 C)    Procedures to minimize patient and personnel exposure,
                       as required by Section 360.40;

                 D)    Use of personnel monitoring devices (if such devices
                       are used at the facility);

                 E)    Film processing procedures; and

                 F)    Prohibited uses of fluoroscopic machines (if such
                       machines are used at the facility), as described in
                       Section 360.40(j).

           2)    The registrant shall maintain for a period of at least 3
                 inspection cycles (see 32 Ill. Adm. Code 410.60(d)),
                 documentation, signed by persons who apply ionizing
                 radiation, that indicates the date and content of training
                 provided.

           3)    The registrant shall provide to each individual subject to
                 in-service training a written policy statement outlining the
                 registrant's radiation safety practices and policies
                 specified in subsection (j)(1) above.

(Source:  Amended at 15 Ill. Reg. 6180, effective April 16, 1991)

Section 360.40  General Equipment and Operation Requirements for Diagnostic
X-ray Systems

The requirements of this Section apply to all diagnostic x-ray systems.
Additional requirements for specific equipment application classes are in
Sections 360.50 through 360.100.

     a)    Half-Value Layer - The half-value layer of the useful beam for a
           given x-ray tube potential shall not be less than the values shown
           in Table B of this Part.  If it is necessary to determine a half-
           value layer at an x-ray tube potential which is not listed in
           Table B, linear interpolation or extrapolation may be utilized to
           determine the appropriate value.

     b)    Beam-On Indicators

           1)    The control panel shall include a device (usually a
                 milliammeter or labeled indicator lamp) which will give
                 positive indication of the production of x-rays whenever the
                 x-ray tube is energized.

           2)    In addition, on certified systems, a signal audible to the
                 operator shall indicate that the exposure has terminated.

     c)    Mechanical Support of Tube Head - The tube housing assembly
           supports shall be adjusted such that the tube housing assembly
           will remain stable during an exposure unless tube housing movement
           is a designed function of the x-ray system.

     d)    Diagnostic Source Assembly Leakage Radiation Limits - The leakage
           radiation measured at a distance of 1 meter from the source shall
           not exceed 100 milliroentgens in 1 hour when the tube is operated
           at its leakage technique factors.

     e)    Exposure Switch - The exposure switch shall be a dead-man switch.

     f)    Patient or Film Support - When a patient or film must be provided
           with auxiliary support during a radiation exposure:

           1)    No person shall be used routinely to hold film or patients;
                 and

           2)    Unless the procedure precludes their use, mechanical holding
                 devices shall be used to restrain patients.  For example,
                 mechanical holding devices could not be used if the devices
                 would preclude clear visualization of the tissue being
                 examined.

                 AGENCY NOTE:  The radiation dose received by radiation
                 workers, patients, and the general public can be reduced if
                 mechanical patient and film support devices are used for
                 radiographic and fluoroscopic procedures.  In the event that
                 an individual must be used in lieu of mechanical patient or
                 film support devices to hold patients or films, every effort
                 should be made to limit the individual's exposure to
                 radiation.  This can be accomplished by not assigning to a
                 single individual the task of supporting patients and films
                 during radiographic and fluoroscopic examinations.  Rather,
                 a number of individuals may be rotated through the
                 assignment, thereby reducing the radiation exposure to one
                 individual.

     g)    Personnel Protection -

           1)    Except for patients who cannot be moved out of the room,
                 only the staff and ancillary personnel required for the
                 medical procedure or training shall be in the room during
                 the radiographic/fluoroscopic exposure.

           2)    Individuals who must be in the room with the patient being
                 radiographed or fluoroscoped shall be positioned such that
                 no part of the individual's body will be exposed to:

                 A)    The useful beam unless protected by 0.5 millimeter
                       lead equivalent apparel or device, and

                 B)    Stray radiation unless protected by 0.25 millimeter
                       lead equivalent apparel or device.

                       AGENCY NOTE:  If apparel is used, it should cover as
                       much of the individual's trunk and upper leg surface
                       areas as possible.  Apparel that protects both
                       posterior and anterior surfaces is recommended.  If a
                       device, e.g., protective screen/barrier, is used in
                       lieu of protective apparel, the device should be of
                       such a width and height to afford protection as would
                       be provided if apparel was worn.

     h)    Technique Guides - In the vicinity of each radiographic x-ray
           system's control panel, a technique guide shall be provided which
           specifies for routine examinations performed with that system, the
           following information:

           AGENCY NOTE:  This requirement is applicable to both dental
           intraoral and extraoral radiographic systems.

           1)    Patient's anatomical size versus technique factors to be
                 utilized,

           2)    Type and size of the film or film-screen combination to be
                 used, and

           3)    SID to be used.

           4)    For automatic exposure control (AEC) systems (i.e., systems
                 employing photo-multiplier tubes or ionization chambers to
                 terminate the x-ray exposure) with selectable exposure
                 detectors and density settings, the technique guide shall
                 also specify the appropriate exposure detector(s) and
                 density setting to be utilized for each radiographic
                 examination listed.

           5)    For AEC systems, the technique guide shall specify the
                 requirements of subsections (h)(1) through (3) above to be
                 followed if operated in a non-automatic mode.

     i)    Patient Exposure Criteria - Procedures and auxiliary equipment
           designed to minimize patient and personnel exposure commensurate
           with needed diagnostic information shall be used.

           AGENCY NOTE:  It is the intent of subsection (i) to provide for
           the optimum optical density on the film while minimizing patient
           exposure.  The kVp and SSD employed in radiographic examinations
           should be as great as practical and consistent with the diagnostic
           objectives of the study.  The x-ray equipment should permit use of
           the optimum kVp that will reduce the doses to the patient based on
           the required optical density of the film.  The milliamperage
           should be high enough to permit as short an exposure time as is
           necessary to limit the effects of motion, which would result in
           the loss of the radiograph's usefulness.  In addition, x-ray
           films, intensifying screens, and other image recording devices
           should be as sensitive as is consistent with the requirements of
           the examination.  Non-screen films should not be used unless
           absolutely necessary for a specific examination.

     j)    Prohibited Use - Fluoroscopy shall not be used as a substitute for
           radiography or in lieu of proper anatomical positioning/centering
           procedures prior to radiographic studies (see 32 Ill. Adm. Code
           310.100 for additional prohibited uses).

     k)    X-ray Film Processing Systems-

           1)    Manual film processing systems shall be monitored by the
                 registrant to assure:

                 A)    The use of a dedicated darkroom timer with an
                       adjustable preset function.  The timer shall be used
                       to adjust film processing time according to solution
                       temperature.

                 B)    The use of a dedicated darkroom thermometer.  The
                       thermometer shall be used to adjust the film
                       processing time according to solution temperature.

                 C)    The use of a film processing guide.  The guide shall
                       contain, at a minimum, information regarding time(s)
                       and temperature(s) (as recommended by the processing
                       chemical manufacturer) used by the registrant to
                       develop radiographs.

                 D)    The frequency at which film processing chemicals are
                       changed.  At a minimum, the interval as recommended by
                       the chemistry manufacturer shall be used.

                 E)    The darkroom safe light illumination is adequate for
                       the film speed(s) and the darkroom operating
                       procedures used to prevent fogging of unprocessed
                       film.

           2)    Automated film processing shall be monitored by the
                 registrant to assure that:

                 A)    The temperature of film processing chemicals is
                       appropriate for the type of film(s) being processed at
                       the film transport speed selected.

                 B)    The film processing chemicals used and their
                       replenishing rate (if applicable) are appropriate for
                       the film transport speed selected.

                 C)    The darkroom safe light illumination is adequate for
                       the film speed(s) and the darkroom operating
                       procedures used (to prevent fogging of unprocessed
                       film).

     l)    Gonadal Shielding - Except for cases in which it would interfere
           with the diagnostic procedure, gonadal shielding of not less than
           0.50 millimeters of lead equivalent shall be used for patients
           (who have not passed the reproductive age) during those
           radiographic procedures in which the gonads are in the useful
           beam.

           AGENCY NOTE:  Protection of the embryo or fetus from radiation
           exposure during radiological examination or treatment of a woman
           of childbearing age (potentially pregnant) should be given special
           consideration.  However, in practice, medical needs should be the
           primary factors in deciding when to administer the examination.

(Source:  Amended at 15 Ill. Reg. 6180, effective April 16, 1991)

Section 360.50  Fluoroscopic Systems

In addition to the provisions of Sections 360.10, 360.30 and 360.40, the
requirements of this Section apply to x-ray equipment and associated
facilities used for fluoroscopy.

     a)    Beam Limitation - The x-ray field shall, whenever possible, be
           limited by adjustable shutters.  In addition:

           1)    For x-ray equipment with adjustable shutters, the
                 mechanism(s) (manual/automatic mode selector(s)) provided
                 for activating and positioning the shutters shall function
                 properly.  This requirement applies to shutters used in both
                 fluoroscopic and spot filming procedures.

           2)    For fluoroscopic equipment without image intensifiers, the
                 x-ray field shall not extend beyond the entire visible area
                 of the image receptor.  This requirement applies to field
                 sizes for both fluoroscopic and spot filming procedures.

                 AGENCY NOTE:  Visible area means that portion of the input
                 surface of the image receptor over which incident x-ray
                 photons are producing an image visible to the fluoroscopist.

           3)    For fluoroscopic equipment with image intensifiers, neither
                 the length nor the width of the x-ray field in the plane of
                 the image receptor shall exceed that of the visible area of
                 the image receptor by more than 3 percent of the SID.  The
                 sum of the excess length and the excess width shall be no
                 greater than 4 percent of the SID.  This requirement applies
                 to field sizes for both fluoroscopic and spot filming
                 procedures.

                 AGENCY NOTE:  Visible area means that portion of the input
                 surface of the image receptor over which incident x-ray
                 photons are producing an image visible to the fluoroscopist.

     b)    Timer - A manual reset, cumulative timing device shall be used
           which will either indicate elapsed on-time by an audible signal or
           turn off the system when the total exposure time exceeds a
           predetermined limit not exceeding 5 minutes in 1 or a series of
           exposures.

     c)    Primary Barrier/Interlock - These devices shall function so that:

           1)    The entire cross section of the useful beam is intercepted
                 by the primary protective barrier of the fluoroscopic image
                 assembly at any SID; and

           2)    The fluoroscopic tube is interlocked to prevent the unit
                 from producing x-rays unless the primary barrier is in
                 position to intercept the useful beam, as specified in
                 subsection (c)(1) above, at all times.

     d)    Source-Skin Distance - The SSD shall not be less than:

           1)    38 centimeters (15 inches) on stationary fluoroscopes which
                 are defined as certified systems,

           2)    35.5 centimeters (14 inches) on stationary fluoroscopes
                 which are defined as non-certified systems,

           3)    30 centimeters (12 inches) on all mobile fluoroscopes,

           4)    20 centimeters (8 inches) for image-intensified fluoroscopes
                 used for a specific surgical application.

     e)    Entrance Exposure Rate (Non-Certified Systems) - Non-certified
           fluoroscopic systems shall not be operable at any combination of
           tube potential and current which will result in an exposure rate
           in excess of 10 Roentgens per minute at the point where the center
           of the useful beam enters the patient.

     f)    Entrance Exposure Rate (Certified Systems) -

           1)    Certified Systems With Automatic Exposure Rate Control -
                 Fluoroscopic systems which are provided with automatic
                 exposure rate control shall not be operable at any
                 combination of tube potential and current which will result
                 in an exposure rate in excess of 10 Roentgens per minute at
                 the point where the center of the useful beam enters the
                 patient, except:

                 A)    During recording of fluoroscopic images; or

                 B)    When an optional high level control is provided.  (See
                       subsection (f)(3) below.)

           2)    Certified Systems Without Automatic Exposure Rate Control -
                 Fluoroscopic systems which are not provided with automatic
                 exposure rate control shall not be operable at any
                 combination of tube potential and current which will result
                 in an exposure rate in excess of 5 Roentgens per minute at
                 the point where the center of the useful beam enters the
                 patient, except:

                 A)    During recording of fluoroscopic images; or

                 B)    When an optional high level control is activated.
                       (See subsection (f)(3) below.)

           3)    When provided with optional high level control, the
                 equipment shall not be operable at any combination of the
                 tube potential and current which will result in an exposure
                 rate in excess of 5 Roentgens per minute at the point where
                 the center of the useful beam enters the patient unless the
                 high level control is activated.

                 A)    Separate means of activation of high level controls
                       shall be required.  The high level control shall only
                       be operable when continuous manual activation is
                       provided by the operator.

                 B)    A continuous signal audible to the fluoroscopist shall
                       indicate that the high level control is being
                       employed.

           4)    Compliance with the requirements of subsections (e) and
                 (f)(1), (2) and (3) above shall be determined as follows:

                 A)    For systems employing automatic exposure rate control,
                       material having an equivalency of at least 0.317 cm
                       (1/8 inch) of lead shall be placed in the primary beam
                       between the image receptor and the radiation measuring
                       device.  The lead or equivalent material shall be
                       positioned to ensure that the entire primary beam is
                       blocked and the radiation measuring device is
                       positioned in accordance with the appropriate
                       measurement protocol outlined in this subsection.

                 B)    If the source is below the table, the exposure rate
                       shall be measured 1 centimeter above the tabletop or
                       cradle.

                       AGENCY NOTE:  The fluoroscopic exposure rate may be
                       measured at a reference point, in the central ray of
                       the primary beam, other than specified in this
                       subsection.  However, compliance shall be determined
                       by calculating the exposure rate at the point
                       specified in this subsection.

                 C)    If the source is above the table, the exposure rate
                       shall be measured at 30 centimeters (12 inches) above
                       the tabletop with the end of the beam-limiting device
                       or spacer positioned as closely as possible to the
                       point of measurement.

                       AGENCY NOTE:  The fluoroscopic exposure rate may be
                       measured at a reference point, in the central ray of
                       the primary beam, other than specified in this
                       subsection.  However, compliance shall be determined
                       by calculating the exposure rate at the point
                       specified in this subsection.

                 D)    For a fixed SID C-arm type of fluoroscope, the
                       exposure rate shall be measured 30 centimeters (12
                       inches) from the input surface of the fluoroscopic
                       imaging assembly.

                       AGENCY NOTE:  The fluoroscopic exposure rate may be
                       measured at a reference point, in the central ray of
                       the primary beam, other than specified in this
                       subsection.  However, compliance shall be determined
                       by calculating the exposure rate at the point
                       specified in this subsection.

                 E)    For a variable SID C-arm type of fluoroscope, the
                       exposure rate shall be measured 30 centimeters (12
                       inches) from the input surface of the fluoroscopic
                       imaging assembly with the end of the beam-limiting
                       device or spacer positioned as close as possible to
                       the point of measurement.

                       AGENCY NOTE:  The fluoroscopic exposure rate may be
                       measured at a reference point, in the central ray of
                       the primary beam, other than specified in this
                       subsection.  However, compliance shall be determined
                       by calculating the exposure rate at the point
                       specified in this subsection.

                 F)    For a lateral type fluoroscope, the exposure rate
                       shall be measured on the central axis of the primary
                       beam at a point 15 centimeters (6 inches) from the
                       centerline of the x-ray table and in the direction of
                       the x-ray source with the end of the beam-limiting
                       device or spacer positioned as closely as possible to
                       the point of measurement.  If the tabletop is movable,
                       it shall be positioned as closely as possible to the
                       lateral x-ray source, with the end of the beam-
                       limiting device or spacer no closer than 15
                       centimeters to the centerline of the x-ray table.

                       AGENCY NOTE:  The fluoroscopic exposure rate may be
                       measured at a reference point, in the central ray of
                       the primary beam, other than specified in this
                       subsection.  However, compliance shall be determined
                       by calculating the exposure rate at the point
                       specified in this subsection.

                       AGENCY NOTE:  A lateral type fluoroscope is a
                       fluoroscope that cannot be rotated so that the source
                       or the fluoroscopic imaging assembly can be positioned
                       below the fluoroscopic table or cradle.

     g)    Screen Shielding - For systems without image intensifiers, the
           fluoroscopic screen shall be covered with transparent protective
           material having a lead equivalency of at least 1.5 millimeters for
           equipment capable of operating up to 100 kVp, at least 1.8
           millimeters for equipment whose maximum operating potential is
           greater than 100 kVp and less than 125 kVp, and at least 2.0
           millimeters for equipment whose maximum operating potential is 125
           kVp or greater.

     h)    Staff and Ancillary Personnel Protection - The fluoroscopist,
           assistants and observers allowed in the examining room shall be
           protected from scatter radiation by protective aprons of not less
           than 0.25 millimeters lead equivalent or whole body protective
           barriers.

     i)    Additional Shielding - A shield of at least 0.25 millimeters lead
           equivalent, such as overlapping protective drapes or hinged or
           sliding panels, should be provided to intercept scatter radiation
           which would otherwise reach the fluoroscopist and others near the
           machine.  This shielding shall not be a substitute for the wearing
           of a protective apron (0.25 mm lead equivalent for protection
           against scattered radiation).

     j)    Mobile Fluoroscopes - In addition to the other requirements of
           this Section, mobile fluoroscopes shall provide intensified
           imaging.

     k)    Radiation Therapy Simulation Systems - Radiation therapy
           simulation systems shall be exempt from the requirements of
           subsections (a), (b), (e) and (f) above provided that:

           1)    Such systems are designed and used in such a manner that no
                 individual other than the patient is in the x-ray room
                 during periods of time when the system is producing x-rays;
                 and

           2)    Such systems that do not meet the requirements of subsection
                 (b) above are provided with a means of indicating the
                 cumulative time that an individual patient has been exposed
                 to x-rays.  Procedures shall require in such cases that the
                 timer be reset between examinations.

     l)    Operator Restrictions - No person shall intentionally administer
           radiation to a human being with a fluoroscopic radiation machine
           unless such person is licensed to practice a treatment of human
           ailments under the Medical Practice Act of 1987, the Illinois
           Dental Practice Act, the Podiatric Medical Practice Act of 1987,
           and the Veterinary Medicine and Surgery Practice Act of 1983,
           except an accredited medical radiographer may operate a
           fluoroscope for static functions when interpretation of the
           results is not required and only under the direct supervision of a
           radiologist who is physically present.

(Source:  Amended at 13 Ill. Reg. 803, effective April 1, 1989)

Section 360.60  Stationary Radiographic Systems Other Than Fluoroscopic,
Dental Intraoral,  Veterinary, or Systems Used Solely For Mammography

In addition to the provisions of Sections 360.10, 360.30 and 360.40, the
requirements of this Section apply to x-ray equipment and associated
facilities used for radiography with stationary radiographic systems other
than fluoroscopic, dental intraoral,  veterinary medical systems, or systems
used solely for mammography.

     a)    Beam Limitation - The useful beam shall be limited to the area of
           clinical interest.  The size of the image receptor utilized for
           each radiographic projection shall be consistent with the
           objectives of the examination.

           1)    Stationary General Purpose X-ray Systems - Means shall be
                 provided to limit the x-ray field in the plane of the image
                 receptor so that such field does not exceed each dimension
                 of the image receptor by more than 2 percent of the SID when
                 the axis of the x-ray beam is perpendicular to the plane of
                 the image receptor.

                 A)    Variable X-Ray Field Limitation - There shall be
                       provided a means for stepless adjustment of the size
                       of the x-ray field.

                 B)    Visual Indication of Field Size - Means shall be
                       provided for visually defining the perimeter of the x-
                       ray field.  The total misalignment of the edges of the
                       visually defined field, with respect to the edges of
                       the x-ray field, shall not exceed 2 percent of the
                       distance from the source to the center of the visually
                       defined field when the surface upon which it appears
                       is perpendicular to the axis of the x-ray beam.

                 C)    Numerical Indication of Field Size -

                       i)    The beam-limiting device shall numerically
                             indicate the x-ray field size in the plane of
                             the image receptor to which it is adjusted.

                       ii)   The x-ray field dimensions shall be specified in
                             inches and/or centimeters, and shall be such
                             that aperture adjustments result in x-ray field
                             dimensions in the plane of the image receptor
                             that do not differ from the numerical indicated
                             dimensions by more than + or - 2 percent of the
                             SID when the beam axis is perpendicular to the
                             plane of the image receptor.

                       iii)  The beam-limiting device shall be provided with
                             SID scales that reflect the actual SID(s) used
                             for radiographic procedures.

                 D)    SID Indication -

                       i)    Means shall be provided to indicate the SID.

                       ii)   SIDs shall be indicated in inches and/or
                             centimeters and the measured SID shall
                             correspond to the indicated value to within 2
                             percent.

                 E)    X-Ray Field/Image Receptor Alignment - Means shall be
                       provided to:

                       i)    Indicate when the axis of the x-ray field is
                             perpendicular to the plane of the image
                             receptor; and

                       ii)   Align the center of the x-ray field with respect
                             to the center of the image receptor to within 2
                             percent of the SID.

                 F)    Additional Requirements for Systems Equipped with
                       Positive Beam Limitation -

                       i)    The x-ray field size in the plane of the image
                             receptor, whether automatically or manually
                             adjusted, shall be such that neither the length
                             nor the width of the x-ray field differs from
                             that of the image receptor by greater than 3
                             percent of the SID and that the sum of the
                             length and width differences without regard to
                             sign be no greater than 4 percent of the SID
                             when the equipment indicates that the beam axis
                             is perpendicular to the plane of the image
                             receptor.

                       ii)   The radiographic system shall be capable of
                             operation, at the discretion of the operator,
                             such that the field size at the image receptor
                             can be adjusted to a size smaller than the image
                             receptor.  The minimum field size at a distance
                             of 100 centimeters (40 inches) shall be equal to
                             or less than 5 centimeters by 5 centimeters (2
                             inches by 2 inches).  Return to positive beam
                             limitation shall occur upon a change in image
                             receptor.

                       iii)  Positive beam limitation may be bypassed when
                             radiography is conducted which does not use the
                             cassette tray or permanently mounted vertical
                             cassette holder, or when either the beam axis or
                             table angulation is not within 10 degrees of the
                             horizontal or vertical during any part of the
                             exposure, or during stereoscopic radiography.
                             If the bypass mode is provided, return to
                             positive beam limitation shall be automatic.

                       iv)   A capability may be provided for overriding
                             positive beam limitation in the event of system
                             failure or to perform special procedures which
                             cannot be performed in the positive mode.  If so
                             provided, a key shall be required to override
                             the positive mode.  It shall be impossible to
                             remove the key while the positive mode is
                             overridden.

           2)    Special Purpose X-Ray Systems -

                 A)    SID Indication -

                       i)    Means shall be provided to indicate the SID.

                       ii)   SIDs shall be indicated in inches and/or
                             centimeters and the measured SID shall
                             correspond to the indicated value to within 2
                             percent.

                 B)    Means shall be provided to limit the x-ray field in
                       the plane of the image receptor so that such field
                       does not exceed each dimension of the image receptor
                       by more than 2 percent of the SID when the axis of the
                       x-ray beam is perpendicular to the plane of the image
                       receptor.

                 C)    Means shall be provided to align the center of the x-
                       ray field with the center of the image receptor to
                       within 2 percent of the SID.

                 D)    The requirements of subsection (a)(2)(B) above may be
                       met:

                       i)    With a system that meets the requirements
                             specified in subsection (a)(1) above; or

                       ii)   With an assortment of removable, fixed-aperture,
                             beam-limiting devices sufficient to meet the
                             requirement for each combination of image
                             receptor size and SID for which the unit is
                             used, with each such device having permanent,
                             clearly legible markings, in inches and/or
                             centimeters, to indicate the image receptor size
                             and SID for which it is designed; or

                       iii)  With a beam-limiting device having multiple
                             fixed apertures sufficient to meet the
                             requirement for each combination of image
                             receptor size and SID for which the unit is
                             used.  Permanent, clearly legible markings, in
                             inches and/or centimeters, shall indicate the
                             image receptor size and SID for which each
                             aperture is designed and shall indicate which
                             aperture is in position for use.

           3)    Single Purpose X-Ray Systems Designed for One Image Receptor
                 Size - Radiographic equipment designed for only 1 image
                 receptor size at a fixed SID shall be provided with means to
                 limit the x-ray field at the plane of the image receptor to
                 dimensions no greater than those of the image receptor when
                 the axis of the x-ray beam is perpendicular to the plane of
                 the image receptor.

     b)    Timers -

           1)    Means shall be provided to terminate the exposure at a
                 preset time interval, preset product of current and time,
                 preset number of pulses, or preset radiation exposure to the
                 image receptor.  Also, it shall not be possible to make an
                 exposure when the timer is set to a zero or off position if
                 either position is provided.

           2)    In addition, for certified systems, termination of exposure
                 shall cause automatic resetting of the timer to its initial
                 setting or to zero.

     c)    Operator's Control Station - Stationary protective barriers shall
           be provided for the x-ray operator.

     d)    Exposure Switch Arrangement - The exposure switch shall be
           arranged so that it cannot be operated by a person outside a
           stationary protective barrier.

     e)    Ancillary Personnel Protection - Individuals other than the
           patient whose presence is required in the radiographic room during
           an x-ray examination shall be protected from scatter radiation by
           protective aprons of not less than 0.25 millimeters lead
           equivalent or whole body protective barriers.

     f)    Medical Radiographic Exposure Limits - The exposure measured at
           the table top for the technique used for an average adult patient
           for routine medical radiography will be the following:  (See
           Appendix A for measurement protocol).

           1)    "Abdomen Anterior Posterior (A.P.) View" exposure shall not
                 exceed 500 milliroentgens per radiograph.

           2)    "Lumbar Spine Lateral View" exposure shall not exceed 1400
                 milliroentgens per radiograph.

           3)    "Cervical Spine A.P. View" exposure shall not exceed 150
                 milliroentgens per radiograph.

           4)    "Skull Posterior Anterior (P.A.) View" exposure shall not
                 exceed 400 milliroentgens per radiograph.

                 AGENCY NOTE:  These exposures are maximums.  With careful
                 selection of technique factors, adjustment of film
                 processing systems, and choice of film and film screen
                 combinations, patient exposures can be further reduced.  For
                 example, the following patient exposures should not be
                 exceeded for each of the exams listed:  "Abdomen A.P. View"
                 exposure should not exceed 350 milliroentgens per
                 radiograph; "Lumbar Spine Lateral View" exposure should not
                 exceed 1,000 milliroentgens per radiograph; "Cervical Spine
                 A.P. View" exposure should not exceed 100 milliroentgens per
                 radiograph; and "Skull P.A. View" exposure should not exceed
                 200 milliroentgens per radiograph.

(Source:  Amended at 15 Ill. Reg. 6180, effective April 16, 1991)

Section 360.70  Mobile/Portable Radiographic Systems Other Than Systems Used
Solely For Mammography

In addition to the provisions of Sections 360.10, 360.30 and 360.40, the
requirements of this Section apply to x-ray equipment and associated
facilities used for radiography with medical mobile/portable systems.

     a)    Beam Limitation - The useful beam shall be limited to the area of
           clinical interest.  The size of the image receptor used for each
           radiographic projection shall be consistent with the objectives of
           the examination.

           1)    Limitation Criteria - Means shall be provided to limit the
                 x-ray field in the plane of the image receptor so that the
                 field does not exceed each dimension of the image receptor
                 by more than 2 percent of the SID when the axis of the x-ray
                 beam is perpendicular to the plane of the image receptor.
                 Limitation of the x-ray field for certified x-ray systems
                 shall be accomplished by the means specified in Section
                 360.60(a)(1)(A) and (B).  For non-certified x-ray systems,
                 the x-ray field shall be limited by the means specified in
                 either Section 360.60(a)(1)(A) and (B) or Section
                 360.60(a)(2)(D)(i) and (ii).

           2)    SID Indication -

                 A)    Means shall be provided to indicate the SID.

                 B)    SIDs shall be indicated in inches and/or centimeters
                       and the measured SID shall correspond to the indicated
                       value to within 2 percent.

     b)    Exposure Switch Arrangement -

           1)    The exposure control switch shall be arranged so that the

                 operator can stand at least 6 feet from the patient, the x-
                 ray tube, and well away from the useful beam.

           2)    All individuals operating mobile/portable x-ray systems
                 shall wear protective aprons of not less than 0.25
                 millimeters lead equivalent.

           3)    When a mobile/portable x-ray system is used in 1 location,
                 it shall be considered a stationary system subject to the
                 requirements specified in Section 360.60(c) and (d).

     c)    Source-Skin Distance -

           1)    Non-certified x-ray systems shall not be operable at a SSD
                 of less than 20 centimeters (8 inches).

           2)    Certified x-ray systems shall not be operable at a SSD of
                 less than 30 centimeters (12 inches).

     d)    Timers -

           1)    Means shall be provided to terminate the exposure at a
                 preset time interval, preset product of current and time,
                 preset number of pulses, or preset radiation exposure to the
                 image receptor.  Also, it shall not be possible to make an
                 exposure when the timer is set to a zero or off position if
                 either position is provided.

           2)    In addition, for certified systems, termination of exposure
                 shall cause automatic resetting of the timer to its initial
                 setting or to zero.

     e)    Radiation from Capacitor Energy Storage X-ray Equipment in Standby
           Status - Radiation emitted from the x-ray tube when the exposure
           switch or timer is not activated shall not exceed a rate of 2
           milliroentgens per hour at 5 centimeters (2 inches) from any
           accessible surface of the diagnostic source assembly, with the
           beam-limiting device fully open.

     f)    Ancillary Personnel Protection - Individuals other than the
           patient whose presence is required in the radiographic room during
           an x-ray examination shall be protected from scatter radiation by
           protective aprons of not less than 0.25 millimeters lead
           equivalent or whole body protective barriers.

     g)    Medical radiographic Exposure Limits - Criteria specified in
           Section 360.60(f) apply.

(Source:  Amended at 15 Ill. Reg. 6180, effective April 16, 1991)

Section 360.71  Requirements for Facilities Performing Mammography

In addition to the provisions of Sections 360.30, 360.40 and 32 Ill. Adm. Code
400 and 401, the requirements of this Section apply to mammography systems and
associated facilities used for mammography.

     a)    Physician Supervision - Mammography operations and procedures
           shall be under the supervision of a physician licensed under the
           Medical Practice Act of 1987 (Ill. Rev. Stat. 1989, ch. 111, par.
           4400) to practice medicine in all of its branches.

           AGENCY NOTE:  The individual interpreting clinical images of the
           breast should be a radiologist trained in the imaging modality
           being used and should be certified or eligible for certification
           by either the American Board of Radiology in diagnostic radiology
           or general radiology or the American Osteopathic Board of
           Radiology.  A facility performing mammography should have a
           program that is accredited by the American College of Radiology or
           the Illinois Radiological Society or have a program that is
           comparable.

     b)    Screen-Film Mammography - Screen-film mammography shall only be
           performed with a special purpose radiation machine specifically
           designed for mammography procedures.

     c)    Xeromammography - Xeromammography shall only be performed with a
           radiation machine that has been specifically designed for or
           modified to perform xeromammography.

     d)    Beam Limitation -

           1)    Means shall be provided to limit the useful beam so that the
                 x-ray field at the plane of the image receptor does not
                 extend beyond any edge of the image receptor at any
                 designated source to image receptor distance (SID).
                 However, the x-ray field may extend beyond the edge of the
                 image receptor designed to be adjacent to the chest wall
                 provided it does not extend beyond this edge by more than 2
                 percent of the SID.

           2)    The requirement of subsection (d)(1) above shall be met with
                 a system that performs as prescribed in either Section
                 360.60(a)(1) or Section 360.60(a)(2)(D)(ii).

     e)    Beam Limiting Device Labeling -

           1)    If beam limitation is met with a device as prescribed in
                 Section 360.60(a)(2)(D)(ii), such device shall have
                 permanent clearly legible markings in inches and/or
                 centimeters indicating the image receptor size and SID for
                 which each device is designed.

           2)    If the radiation machine is capable of using variable SIDs,
                 the SID indication specified in subsection (e)(1) above
                 shall be the maximum SID for which the beam limiting device
                 is designed.

     f)    Exposure Switch Arrangement -

           1)    Stationary Mammography Systems - For mammography systems
                 consisting of stationary x-ray equipment, the criteria
                 specified in Section 360.60(d) shall apply.

           2)    Mobile and Portable Mammography Systems - For mammography
                 systems consisting of mobile and portable x-ray equipment,
                 the criteria specified in either Section 360.60(d) or
                 Section 360.70(b)(1) shall apply.

     g)    Operator Shielding -

           1)    Stationary Mammography Systems - For mammography systems
                 consisting of stationary x-ray equipment, the criteria
                 specified in Section 360.60(c) shall apply.

           2)    Mobile and Portable Mammography Systems - For mammography
                 systems consisting of mobile and portable x-ray equipment,
                 the criteria specified in either Section 360.60(c) or
                 Section 360.70(b)(2) shall apply.

           3)    Mobile and Portable X-Ray Equipment Used in Only One
                 Location - When mobile or portable x-ray equipment is used
                 in only one location, it shall be considered a stationary
                 system and shall be subject to the requirements specified in
                 Sections 360.60(c) and (d).

     h)    Timers - The criteria specified in Section 360.60(b) shall apply.

     i)    SID Indicator - For radiation machines capable of operating at
           variable SIDs, the criteria specified in Section 360.60(a)(1)(D)
           shall apply.

     j)    X-ray Field/Image Receptor Alignment - Section 360.60(a)(2)(C)
           shall apply.

     k)    Backup Timer - Radiation systems with automatic exposure control
           (AEC) shall incorporate a backup timer to terminate the radiation
           exposure in the event of AEC failure.

     l)    SSD - Except with respect to magnification for mammography
           projections, mammography systems shall satisfy the requirements of
           Section 360.70(c).

     m)    Focal Spot Size - The nominal focal spot size, as specified by the
           x-ray tube manufacturer, shall not exceed 0.7 millimeters.

     n)    Reproducibility of Exposures - For all systems, regardless of
           whether they are equipped with AEC, the estimated coefficient of
           variation of radiation exposures shall be no greater than 0.05 for
           any specific combination of selected technique factors.  It will
           not be necessary to calculate the coefficient of variation if for
           4 consecutive measurements taken within a time period of 1 hour
           the difference between the highest and lowest exposures does not
           exceed 10 percent. If there is a difference greater than 10
           percent, 6 additional exposures shall be made and the coefficient
           of variation shall be calculated.

           AGENCY NOTE:  "Coefficient of variation" means the ratio of the
           standard deviation to the mean value of a population of
           observations.  It is calculated using the following equation:








           where:

           s = Estimated standard deviation of the population.

           X = Mean value of observations in sample.

           Xi = ith observation sampled.

           n = Number of observations sampled.

     o)    Mammography Exam Dose Limits (See Appendix B for the required
           measurement protocol.) - The mean glandular dose for one
           craniocaudal view of a 1.8 inch (4.5 centimeter) compressed breast
           (50 percent adipose/50 percent glandular) shall not exceed:

           1)    100 millirads (1 milligray) for film/screen radiographs not
                 employing the use of grids,

           2)    300 millirads (3 milligrays) for film/screen radiographs
                 employing the use of grids, or

           3)    400 millirads (4 milligrays) for xerography.

                 AGENCY NOTE:  A facility performing mammography should have
                 a Mammography Dosimetry Test phantom specifically designed
                 for breast imaging.  This phantom should enable the
                 individual interpreting clinical images of the breast to
                 readily determine the x-ray system's ability to identify
                 masses, fibrils, and calcifications on a month to month
                 basis.

(Source:  Added at 15 Ill. Reg. 6180, effective April 16, 1991)

Section 360.80  Photofluorographic Systems

In addition to the provisions of Sections 360.10, 360.30 and 360.40, the
requirements of this Section apply to x-ray equipment and associated
facilities used for photofluorography.

AGENCY NOTE:  Photofluorography is frequently called mass miniature
radiography.  This is the method usually employed in projects such as chest x-
rays for tuberculosis control.  In this technique the image of a fluorescent
screen is recorded on film by means of a camera.  The film used is of small
size.  Some units use cut film 4 inches by 4 inches in size.  More commonly,
roll film is used; the film is usually 70-mm wide, but may be 90 or 100-mm.

     a)    Beam Limitation - Photofluorographic systems shall be provided
           with means to limit the x-ray field at the plane of the image
           receptor to dimensions no greater than those of the image
           receptor.

     b)    Operator Control Station - Criteria specified in Section 360.60(c)
           shall apply.

     c)    Exposure Switch Arrangement - Criteria Specified in Section
           360.60(d) shall apply.

     d)    Timers -

           1)    Means shall be provided to terminate the exposure at a
                 preset time interval, preset product of current and time,
                 preset number of pulses, or preset radiation exposure to the
                 image receptor.  Also, it shall not be possible to make an
                 exposure when the timer is set to a zero or off position if
                 either position is provided.

           2)    In addition, for certified systems, termination of exposure
                 shall cause automatic resetting of the timer to its initial
                 setting or to zero.

     e)    Ancillary Personnel Protection - Other than the patient,
           individuals whose presence is required in the radiographic room
           during an x-ray examination shall be protected from scatter
           radiation by protective aprons of not less than 0.25 millimeters
           lead equivalent or whole body protective barriers.

     f)    Photofluorographic Exposure Limit - The exposure to an average
           patient shall not exceed 200 milliroentgens per photofluorograph
           (See Appendix A for measurement protocol).

           AGENCY NOTE:  The patient exposure for this procedure should not
           exceed 100 milliroentgens per photofluorograph.

     g)    Medical Supervision - The supervising physician shall outline
           responsibilities regarding the photofluorographic operating and
           patient screening procedures.  The procedures shall be submitted
           to this Department in writing prior to utilization of the
           equipment.  Operating and patient screening procedures shall
           include at a minimum the following:

           1)    Operator qualifications,

           2)    Operator supervision,

           3)    Methods of operating x-ray machines,

           4)    Patient age limit,

           5)    Frequency of exam,

           6)    Pregnancy cases, etc.

(Source:  Amended at 13 Ill. Reg. 803, effective April 1, 1989)

Section 360.90  Intraoral Dental Radiographic Systems

In addition to the provisions of Sections 360.10, 360.30 and 360.40, the
requirements of this Section apply to x-ray equipment and associated
facilities used for intraoral dental radiography.

     a)    Beam Limitation - X-ray systems designed for use with an intraoral
           image receptor shall be provided with means to limit the x-ray
           field at the patient's face to the smallest area which is
           clinically necessary.  The x-ray field striking the patient's face
           shall not exceed a circle 7.5 centimeters (3 inches) in diameter.
           Even though the beam shall be containable within a circle 7.5
           centimeters in diameter, it may be a rectangular configuration.

           1)    Beam limitation shall be accomplished by a position
                 indicating device that is operated and shielded.  The device
                 shall provide the same degree of shielding as the tube
                 housing assembly.

           2)    The position indicating device shall provide a SSD of not
                 less than 18 centimeters (7 inches) for systems operated
                 above fifty (50) kVp or 10 centimeters (4 inches) for
                 systems operated at fifty (50) kVp or below.

     b)    Timers -

           1)    Means shall be provided to terminate the exposure at a
                 preset time interval, preset product of current and time,
                 preset number of pulses, or preset radiation exposure to the
                 image receptor.  Also, it shall not be possible to make an
                 exposure when the timer is set to a zero or off position if
                 either position is provided.

           2)    In addition, for certified systems, termination of exposure
                 shall cause automatic resetting of the timer to its initial
                 setting or to zero.

     c)    Exposure Switch Arrangement/Operator Protection - The exposure
           switch shall be arranged so that the operator can be at least 6
           feet from the patient and well away from the useful beam during an
           exposure.

           AGENCY NOTE:  The Department recommends that whenever practicable,
           the x-ray operator stand behind a protective barrier or be
           provided with a protective apron of not less than 0.25 millimeters
           lead equivalent.

     d)    Ancillary Personnel Protection - Other than the operator and the
           patient, individuals whose presence is required in the room during
           an x-ray examination shall be protected from stray radiation by
           protective aprons of not less than 0.25 millimeters lead
           equivalent or a protective barrier.

     e)    Dental Radiographic Exposure Limits (Single Film) - The entrance
           exposure to an adult patient for a routine intraoral bitewing exam
           shall not exceed the maximum limits specified in Table C.

(Source:  Amended at 13 Ill. Reg. 803, effective April 1, 1989)

Section 360.100  Veterinary Radiographic Systems

In addition to the provisions of Sections 360.10, 360.30 and 360.40, the
requirements of this Section apply to x-ray equipment and associated
facilities used for radiography with veterinary systems.

     a)    Beam Limitation - The useful beam shall be limited to the area of
           clinical interest.  The size of the image receptor used for each
           radiographic projection shall be consistent with the objectives of
           the examination.

           1)    Limitation Criteria - Means shall be provided to limit the
                 x-ray field in the plane of the image receptor so that the
                 field does not exceed each dimension of the image receptor
                 by more than 2 percent of the SID when the axis of the x-ray
                 beam is perpendicular to the plane of the image receptor.

           2)    Means shall be provided to align the center of the x-ray
                 field with the center of the image receptor to within 2
                 percent of the SID.

           3)    The requirements of subsection (a)(1) above may be met with:

                 A)    A system that meets the requirements specified in
                       Section 360.60(a)(1); or

                 B)    An assortment of removable, fixed-aperture, beam-
                       limiting devices sufficient to meet the requirement
                       for each combination of image receptor size and SID
                       for which the unit is used, with each such device
                       having permanent, clearly legible markings in inches

                       and/or centimeters, to indicate the image receptor
                       size and SID for which it is designed; or

                 C)    A beam-limiting device having multiple fixed apertures
                       sufficient to meet the requirement for each
                       combination of image receptor size and SID for which
                       the unit is used.  Permanent, clearly legible
                       markings, in inches and/or centimeters, shall indicate
                       the image receptor size and SID for which each
                       aperture is designed and shall indicate which aperture
                       is in position for use.

           4)    SID Indication -

                 A)    Means shall be provided to indicate the SID.

                 B)    SIDs shall be indicated in inches and/or centimeters
                       and the measured SID shall correspond to the indicated
                       value to within 2 percent.

     b)    Exposure Switch Arrangement -

           1)    The exposure control switch shall be arranged so the
                 operator can be at least 6 feet from the patient, the x-ray
                 tube, and well away from the useful beam.

           2)    All individuals operating veterinary x-ray systems shall
                 wear protective aprons of not less than 0.25 millimeters
                 lead equivalent or shall be protected from scatter radiation
                 by a protective barrier.

     c)    Timers -

           1)    Means shall be provided to terminate the exposure at a
                 preset time interval, preset product of current and time,
                 preset number of pulses, or preset radiation exposure to the
                 image receptor.  Also, it shall not be possible to make an
                 exposure when the timer is set to a zero or off position if
                 either position is provided.

           2)    In addition, for certified systems, termination of exposure
                 shall cause automatic resetting of the timer to its initial
                 setting or to zero.

     d)    Radiation From Capacitor Energy Storage X-ray Equipment in Standby
           Status - Radiation emitted from the x-ray tube when the exposure
           switch or timer is not activated shall not exceed a rate of 2
           milliroentgens per hour at 5 centimeters from any accessible
           surface of the diagnostic source assembly, with the beam-limiting
           device fully open.

     e)    Ancillary Personnel Protection - Individuals whose presence is
           required during an x-ray examination shall be protected from
           scatter radiation by protective aprons or gowns of not less than
           0.25 millimeters lead equivalent or whole body protective
           barriers.

     f)    Veterinary Fluoroscopic Systems - All provisions of Section 360.50
           apply except subsections (b) and (l).

(Source:  Amended at 13 Ill. Reg. 803, effective April 1, 1989)

Section 360.110  Therapeutic X-Ray Installations

     a)    Equipment

           1)    Tube Housing -

                 A)    The protective tube housing shall be of therapeutic-
                       type, except in the case of "Contact" X-Ray Therapy
                       (See Section 360.120(a)).

                 B)    A therapeutic-type protective tube/target housing is
                       defined as follows:

                       i)    For x-ray therapy equipment not capable of
                             operating at 500 kVp or above, the following
                             definition applies:  An x-ray tube housing so
                             constructed that the leakage radiation at a
                             distance of one meter from the source does not
                             exceed one Roentgen in an hour when the tube is
                             operated at its maximum rated continuous current
                             for the maximum rated tube potential.

                       ii)   For x-ray therapy equipment capable of operating
                             at 500 kVp or above, the following definition
                             applies:  An x-ray tube housing so constructed
                             that the leakage radiation at a distance of one
                             meter from the source does not exceed either one
                             Roentgen in an hour or 0.1 percent of the useful
                             beam dose rate at one meter from the source,
                             whichever is the greater when the machine is
                             operated at its maximum rated continuous current
                             for the maximum rated accelerating potential.

                       iii)  In either case, small areas of reduced
                             protection are acceptable provided the average
                             reading over any 100 square centimeters area at
                             one meter distance from the source does not
                             exceed the values given above.

           2)    Collimation - The useful beam shall be collimated by means
                 of cones, diaphragms or lead coverings to reduce the useful
                 beam striking a patient to the smallest area which is
                 clinically necessary.

           3)    Filtration - The filter system shall conform to recommended
                 practices as defined in the current National Council on
                 Radiation Protection Report No. 33, entitled "Medical X-Ray
                 and Gamma Ray Protection for Energies up to 10 MeV,
                 Equipment Design and Use".

           4)    Filtration Indicator - A filter indication system shall be
                 used on all therapy machines using changeable filters.  It
                 shall be designed so as to permit easy recognition of any
                 added filter in place.  It shall indicate, from the control
                 panel, the presence or absence of any filter.

           5)    Tube/Aperture Alignment - The x-ray tube shall be so mounted
                 that it cannot turn or slide with respect to the housing
                 aperture.

           6)    Tube Head Stability - Methods shall be provided to
                 immobilize the tube housing during stationary portal
                 treatment.

           7)    Preset Exposure Termination Device - A device (e.g. an
                 automatic timer, exposure meter or dose meter) shall be
                 provided to terminate the exposure after a preset time
                 interval or preset exposure or dose limit.  Methods shall be
                 provided for the operator to terminate the exposure at any
                 time.

           8)    Shutter Position Indicator - Equipment utilizing shutters to
                 control the useful beam shall have a shutter position
                 indicator on the control panel.

           9)    Positive Indicator - The control panel shall include a
                 device (usually a milliammeter) which will give positive
                 indication of the production of x-rays whenever the x-ray
                 tube is energized.

           10)   Interlock - 150 kVp or Greater - A reliable interlock shall
                 be installed on each door of the therapeutic room.  This
                 interlock shall be wired into the electrical circuit in such
                 a manner that when the door is opened, for any reason, the
                 x-ray exposure will automatically be terminated so that
                 irradiation can be resumed only by manually resetting the
                 controls on the control panel after the door is closed.

           11)   Door-Fastening Mechanism - If required, the door-fastening
                 mechanism shall be installed so that the door can be opened
                 from the inside in case of emergency.

           12)   Warning Lights - Treatment rooms to which access is possible
                 through more than one entrance shall be provided with
                 flashing warning lights in a readily observable position
                 near the outside of all access doors, which will indicate
                 when the useful beam is "on".

     b)    Operating Procedures.

           1)    All new facilities, and existing facilities not previously
                 surveyed, shall have a protection survey made by, or under
                 the direction of a qualified expert.  This shall also be
                 done after any change in the facility which might produce a
                 radiation hazard.  The expert shall report his findings in
                 writing to the person in charge of the facility and a copy
                 of this report shall be transmitted to the Department.

           2)    The facility shall be operated in compliance with any
                 limitations indicated by the protection survey required by
                 the Department.

           3)    When a patient must be held in position for radiation
                 therapy, mechanical supporting or restraining devices shall
                 be used whenever feasible.

           4)    Attendants - Below 120 kVp - No person, other than the
                 attending physician or his assistant, shall be in the
                 treatment room during exposure, except if in the opinion of
                 the physician such attendance in the room is clinically
                 necessary or desirable.  If a person is required to hold or
                 support the patient, no portion of such person shall be in
                 the useful beam and such person shall be protected as much
                 as is practicable from scattered radiation.  The exposure of
                 any person used for this purpose shall be monitored and a
                 record maintained.

           5)    Attendants - 120 kVp or Greater - No person except the
                 attending physician who works with ionizing radiation, shall
                 be in the treatment room during exposure.

           6)    Communication with Patient - The control panel shall be
                 arranged in such a manner that the operator can maintain
                 aural and visual communication with the patient.  Both the
                 patient and the control panel shall be under observation of
                 the operator during the entire exposure.

           7)    The output of each therapeutic x-ray machine shall be
                 calibrated by, or under the direction of a qualified expert.
                 The calibration shall be repeated after any change in or
                 replacement of components of the x-ray output.  Check
                 calibrations shall be made at least once a year thereafter
                 and be maintained by the registrant.  Records of calibration
                 shall be posted at the controls and shall include the date
                 of calibration and the written signature of the qualified
                 expert.

     c)    Structural Shielding.

           1)    Primary protective barriers shall be provided for any area
                 that the useful beam may strike when using the largest
                 possible diaphragm opening.  Such barriers should extend at
                 least one foot beyond the useful beam for any possible
                 orientation.*

           2)    Secondary protective barriers shall be provided for all
                 occupied areas exposed to leakage and scattered radiation.*

                 *Shall also conform with criteria outlined in NCRP Report
                 No. 49 entitled "Structural Shielding Design and Evaluation
                 for Medical Use of X-Ray and Gamma Rays of Energies Up to 10
                 MeV."

Section 360.120  Special Requirements for X-Ray Therapy Equipment Operated at
Potential of Fifty (50) kVp and Below

     a)    Equipment.

           1)    All provisions of Section 360.110(a) apply except
                 paragraphs(a)(10),(11) and (12), and except in the case of
                 "Contact" x-ray tubes, where the leakage radiation at the
                 surface of the housing shall not exceed 0.1 R/hr.

           2)    Preset Timer - Automatic timers shall be provided which will
                 permit accurate presetting and termination of exposures as
                 short as one second.

     b)    Operating Procedures.

           1)    In the therapeutic application of apparatus constructed with
                 beryllium or other low-filtration windows, the registrant
                 shall insure that the unfiltered radiation reaches only the
                 part intended and that the useful beam is blocked at all
                 time except when actually being used.

           2)    Machines having an output of more than 1,000 Roentgens per
                 minute at any accessible place shall not be left unattended
                 without the power being shut off at the main disconnect
                 switch in addition to the control panel switch.

           3)    If the tube must be hand-held during irradiation the
                 operator shall wear protective gloves and protective apron
                 no less than 0.5 millimeter lead equivalent.

     c)    Structural Shielding.

           In general, additional structural barriers will not be required.
           However, calculations of dose rates at critical positions beyond
           walls, partitions, etc., shall be made to determine if additional
           structural barriers will be required.


Section 360.APPENDIX A  Medical Radiographic Exposure Limits

                            Measurement Protocol

The medical radiographic examinations specified in Section 360.60(f) are
examinations frequently taken at medical radiation installations.
The exposures measured using the "measurement protocol" are not actual
"entrance exposures" to patients but should be considered "reference
exposures" (patients are not involved in the measurement protocol).

     a)    An integrating exposure radiation measuring device is placed on
           the radiographic table-top directly in the center of the useful
           beam.

           AGENCY NOTE:  Applicable radiation measuring devices include but
           are not limited to:  Condenser - R meter, low energy dosimeter
           (LED), integrating mode of a cutie-pie, thermoluminescent
           dosimetry (TLD).

           AGENCY NOTE:  In the event a patient support table is not
           utilized, (e.g., mobile/portable radiography) the measuring device
           should be placed directly in the center of the useful beam on any
           convenient support surface.  The radiographic tube-to-measuring
           device distance shall be identical to the source-to-image receptor
           distance used during an actual patient exposure.

     b)    The radiographic tube is positioned identical to that used during
           an actual patient exposure (usually 40 inches to bucky).

     c)    The exposure technique is set up as follows:

           1)    For non-phototimed x-ray systems, the exposure technique
                 used by the x-ray operator for an "average-sized" adult
                 patient is set up on the controls.

                 EXAMPLE:  A.P. Abdomen (80 kVp, 100 mAs).  The useful beam
                 is limited to the appropriate film size.

           2)    For phototimed x-ray systems, including photofluorographic
                 systems, the exposure technique used by the x-ray operator
                 for an "average sized" adult patient is set up on the
                 controls.  An appropriate phantom (simulating body
                 attenuation) is placed in the useful beam between the
                 radiation measuring device and the radiographic table-top or
                 the surface of the photofluorographic screen assembly.  The
                 useful beam is limited to the size of the phantom.

     d)    A radiographic exposure is made (without patient) and the reading
           obtained from the radiation measuring device is recorded as a
           representative radiation exposure for that specific examination to
           determine compliance.

(Source:  Amended at 13 Ill. Reg. 803, effective April 1, 1989)
Section 360.APPENDIX B  Mammography Dose Limit

                            MEASUREMENT PROTOCOL

The technique factors used for performing a mammography examination shall not
permit the mean glandular absorbed dose to exceed the limits specified in
Section 360.71(o).

The mammography exam dose limits are based on an average compressed breast
value of 4.5 centimeters having an average density (i.e., 50 percent adipose
and 50 percent glandular).  While other sizes and densities may be present in
the actual population of interest, this average compressed breast of 4.5
centimeters and density of 50 percent adipose and 50 percent glandular tissue
will suffice as a dose model reference within the accuracies needed to provide
safety to the general public.


     a)    Non-Automatic Exposure Control Systems -  Perform the following
           steps to determine the mean glandular dose to a nominal 4.5
           centimeter compressed breast:

           1)    Determine the x-ray tube target material.

           2)    Identify the curve which represents a 4.5 centimeter
                 compressed breast thickness from the "Mammography Dose
                 Evaluation" graph (see Illustration B).

           3)    Measure and record the x-ray system's useful beam half-value
                 layer (HVL). (See Section 360.40(a).)  Any compression
                 device normally in the useful beam during mammography
                 procedures shall be required to be placed between the x-ray
                 tube target and measuring device when determining the HVL.
                 The useful beam shall be collimated to a size encompassing
                 the detector.

                 AGENCY NOTE:  Filters used for the HVL evaluation should be
                 placed as close to the target as possible.  The HVL for
                 film/screen mammography should not exceed 0.4 millimeters of
                 aluminum equivalent and 1.6 millimeters of aluminum
                 equivalent for xerography.

           4)    Estimate the normalized rad per roentgen factor from the
                 "Mammography Dose Evaluation" graph (see Illustration B)
                 using the coordinates for a 4.5 centimeter compressed breast
                 thickness and the appropriate HVL.

                 Example:    A radiation machine is provided with a
                             molybdenum target and its HVL is determined to
                             be 0.3.  Therefore, for a 4.5 centimeter
                             compressed breast, the normalized rad per
                             roentgen factor would be 0.15.

           5)    Set the appropriate craniocaudal source to image receptor
                 distance (SID) for the image receptor system used.

           6)    Place a radiation measuring device in the useful beam so
                 that the center axis of the device is parallel to the breast
                 support assembly (BSA).  The geometric center of the
                 measuring device shall be positioned so that it is centered
                 4.5 centimeters above the BSA, 2.54 centimeters (1 inch)
                 from the chest wall edge of the BSA and at the center line
                 of the BSA.  (see Illustration A) The radiation measuring
                 device shall be an integrating type appropriate to the high
                 beam intensity and mammographic kilovoltage peak (kVp) used.
                 No part of the device's detector area shall be outside of
                 the useful beam.

           7)    Collimate the x-ray field to the size normally used.

           8)    Set the milliamperage (mA), kilovoltage peak (kVp) and
                 exposure time technique factors normally used for a nominal
                 4.5 centimeter compressed breast.

           9)    Position in the useful beam any compression apparatus
                 normally used.

           10)   Measure and record the exposure in air with the radiation
                 measuring device.

           11)   Calculate the mean glandular dose for a 4.5 centimeter
                 compressed breast by multiplying the measured exposure in
                 roentgens by the normalized rads per roentgen figure
                 established in subsection (a)(4) above.

                 Example:    The rad (milligray) per roentgen figure of 0.15
                             established in subsection (a)(4) will be used.
                             The measured roentgen output (subsection (a)(10)
                             is determined to be 1.8 roentgens.  Therefore,
                             the mean glandular dose would be 288 millirads
                             (2.88 milligrays).  If the image receptor type
                             used was film/screen with grid, the system would
                             be in compliance with Section 360.71(o)(2).

     b)    Systems Equipped with AEC That Have A Means To Determine Exposure
           Elapsed Time - Perform one of the following two procedures to
           determine the mean glandular dose to a 4.5 centimeter compressed
           breast for systems equipped with either a milliamp second (mAs)
           meter or an exposure time indicator.

           1)    Mammography dosimetry test phantom procedure

                 A)    Follow the procedures outlined in subsections (a)(1),
                       (2), (3), (4) and (5) above.

                 B)    Place a mammography dosimetry test phantom (see the
                       definition for "Mammography dosimetry test phantom" in
                       Section 360.20) on the breast support assembly.
                       Center the phantom on the assembly to assure the
                       phantom is over the automatic exposure control
                       device(s).

                 C)    Place a radiation measuring device in the useful beam
                       so the center axis of the device is parallel to the
                       breast support assembly (BSA).  The geometric center
                       of the measuring device shall be positioned so that it
                       is centered 4.5 centimeters above the BSA, 2.54
                       centimeters (1 inch) from the chest wall edge of the
                       BSA and immediately adjacent to either side of the
                       breast dosimetry test phantom.

                 D)    Collimate the x-ray field to the size normally used
                       and assure that the area covered by the useful beam
                       includes the breast dosimetry test phantom and the
                       detector area of the radiation measuring device.

                 E)    Set the milliamperage (mA) and kilovoltage peak (kVp)
                       technique factors normally used for a nominal 4.5
                       centimeter compressed breast.

                 F)    Position in the useful beam, any compression apparatus
                       normally used.

                 G)    Make an exposure and record the exposure indicated by
                       the radiation measuring device.

                 H)    Calculate the mean glandular dose for a 4.5 centimeter
                       compressed breast. (See subsection (a)(11) above.)

           2)    Mammography test procedure without a mammography dosimetry
                 test phantom.

                 A)    Record for 30 consecutive days the patient
                       characteristics in terms of mAs, compressed breast
                       size and kVp per radiograph to determine an average.
                       All patients radiographed during the 30 day period
                       shall be included in the average.  For the purpose of
                       this protocol, breast size measurements shall be
                       accurate to within 0.3 centimeters.  The 4.5
                       centimeter compressed dimension is measured at a point
                       midway between the medial and lateral edges of the
                       breast and 1/2 the distance between the surface of the
                       chest wall and the base of the nipple.

                 B)    Calculate the average mAs and kVp associated with a
                       nominal 4.5 centimeter compressed breast from the data
                       recorded.  The calculation may be either arithmetic or
                       graphic.

                       AGENCY NOTE:  A sample size of at least 30 will be
                       considered an acceptable sample size for the purpose
                       of determining the average mAs and kVp used for a
                       nominal 4.5 centimeter compressed breast.  If the
                       sample size is less than 30, the graphic method of
                       determining the average mAs and kVp shall be used.

                 C)    Turn off the automatic exposure control system and
                       set-up the x-ray system as specified in subsections
                       (a)(1), (2), (3), (4), (5) and (6) above.

                 D)    Collimate the x-ray field to the size normally used
                       and assure that the area of the useful beam includes
                       the detector area of the radiation measuring device.

                 E)    Set the mAs and kVp as would be used if the automatic
                       system was being used for a 4.5 centimeter compressed
                       breast (i.e., based on the average technique factors
                       calculated from subsection (b)(2)(B) above).

                 F)    Position in the useful beam any compression apparatus
                       normally used.

                 G)    Perform the exposure measurement and calculate the
                       mean glandular dose as specified in subsections
                       (a)(10) and (11) above.

     c)    Systems Equipped with AEC That Do Not Have A Means To Determine
           Exposure Elapsed Time - Perform one of the following two
           procedures to determine the mean glandular dose to a nominal 4.5
           centimeter compressed breast for systems not equipped with either
           an mAs or elapsed exposure time indicator:

           1)    Procedure No. 1 - No Mammography Dosimetry Test Phantom

                 A)    Temporarily install either an mAs meter or elapsed
                       exposure on-time indicator on the x-ray system.  This
                       installation shall be done by a person competent in
                       electrical circuitry to avoid electrical hazards.

                 B)    Follow the procedures listed under subsection (b)(2)
                       above.

           2)    Procedure No. 2 - With a Dosimetry Test Phantom

                 Follow the procedures in subsection (b)(1) above.


(Source:  Added at 15 Ill. Reg. 6180, effective April 16, 1991)
Section 360.ILLUSTRATION A  Thimble and Pancake Chamber - Radiation Measuring
Device


















(Source:  Added at 15 Ill. Reg. 6180, effective April 16, 1991) Section 360.ILLUSTRATION B  Mammography Dose Evaluation Graph






































Working curves for evaluating the average glandular dose delivered by 1 R in
air incident on a 4.5 centimeter compressed breast thickness (vertical axis)
vs. first HVL in mmAl (horizontal axis).  The solid line represents tungsten
tube target data.  The broken line represents molybdenum or molybdenum/
tungsten target tube data.  (Excerpt from NCRP 85)

(Source:  Added at 15 Ill. Reg. 6180, effective April 16, 1991)
Section 360.TABLE A  Filtration Required as a Function of Operating kVp
(Repealed)


(Source:  Repealed at 13 Ill. Reg. 803, effective April 1, 1989)


Section 360.TABLE B  Half-Value Layer as a Function of Tube Potential

                       X-ray Tube Voltage
                       (kilovolt peak)          Minimum HVL (mm of Al) (1)

                            Indicated           Specified         Other
Designed operating           Operating           Dental            X-Ray
     range                  Potential (2)       Systems (3)       Systems (4)

Below 50                          30                1.5               0.3
                                 40                1.5               0.4
                                 49                1.5               0.5

50 to 70                          50                1.5               1.2
                                 60                1.5               1.3
                                 70                1.5               1.5

Above 71                          71                2.1               2.1
                                 80                2.3               2.3
                                 90                2.5               2.5
                                100                2.7               2.7
                                110                3.0               3.0
                                120                3.2               3.2
                                130                3.5               3.5
                                140                3.8               3.8
                                150                4.1               4.1

(1)  AGENCY NOTE:  Linear extrapolation or interpolation may be made for an
    x-ray tube potential (kVp) not listed in Table B above (e.g., in the
    column entitled "Other X-ray Systems" operated at 20 kVp and 95 kVp, the
    minimum HVL required would be 0.2 and 2.6 mm of Al respectively).

(2)  AGENCY NOTE:  If the HVL determination for an x-ray system is below the
    minimum value specified for a given voltage, as indicated at the control
    panel, the actual kilovoltage should be measured and the HVL
    reevaluated.

(3)  AGENCY NOTE:  "Specified Dental Systems" means any dental x-ray system
    designed for use with intraoral image receptors and manufactured after
    December 1, 1980.

(4)  AGENCY NOTE:  "Other X-ray Systems" means all x-ray systems required to
    meet the provisions of Sections 360.50, 360.60, 360.70, 360.80, 360.90
    (except "Specified Dental Systems") and 360.100.

(Source:  Amended at 13 Ill. Reg. 803, effective April 1, 1989)


Section 360.TABLE C  Entrance Exposure Limits Per Intraoral Bitewing Film

    Operating           Maximum Exposure (1)      Minimum Exposure (2)
      (kVp)             (milliroentgens)             (milliroentgens)

       45                       640                        430

       50                       600                        400

       55                       560                        370

       60                       520                        320

       65                       480                        270

       70                       440                        220

       75                       400                        175

       80                       360                        140

       85                       320                        115

       90                       280                        100

       95                       240                        95

      100                       200                        90

(1)  AGENCY NOTE:  Linear extrapolation or interpolation may be made for an
    x-ray tube potential (kVp)  not listed in Table B above (e.g., bitewing
    radiographs taken at 44 kVp and 72 kVp, the maximum entrance exposure
    permitted would be 648 milliroentgens and 424 milliroentgens
    respectively).

(2)  AGENCY NOTE:  The minimum exposures specified in the above table are
    included as recommendations only.  They were empirically determined by a
    panel of dentists in a U.S. FDA, BRH study.  They represent the minimum
    exposure which was found to be necessary to produce a diagnostic quality
    radiograph when a dental phantom, speed group "D" film, and adequate
    film development procedures were used.  However, some x-ray units
    manufactured after 1980, or x-ray units used in conjunction with dental
    film of the ultra speed group "E", may be capable of generating
    exposures lower than listed in this table.

(Source:  Amended at 15 Ill. Reg. 6180, effective April 16, 1991)